STORAGE & CLEANING

OS418L should be stored at room temperature. The sensor lens is the most delicate part of the thermometer. The lens
should be kept clean at all times, care should be taken when cleaning the lens using only a soft cloth or cotton swab

with water or medical alcohol, allowing the lens to fully dry before using the thermometer. Do not submerge any part of

the thermometer.

LCD ERROR MESSAGES

The thermometer incorporates visual diagnostic messages as follows:

‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the settings of
HAL and LAL.

_

_ _

_

‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient

temperature. ‘Er3’ is displayed when the ambient temperature exceeds 0

o

C (32°F) or +50

o

C

(122°F). The thermometer should be allowed plenty of time (minimum 30 minutes) to
stabilize to the working/room temperature.

_

_

Error 5~9

, for all other error messages it is necessary to reset the thermometer. To reset it,

turn the instrument off, remove the battery and wait for a minimum of one minute, reinsert
the battery and turn 

_

__

_

‘Hi’  or  ’Lo’ is displayed when the temperature being measured is outside of the
measurement range.

BATTERIES

The thermometer incorporates visual low battery indication as follows:

‘Battery OK’: measurements

 are possible

‘Battery Low’: battery needs to be

replaced, measurements are still possible

‘Battery Exhausted’: measurements are

not possible

 When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries.

Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.

 Dispose of used battery promptly and keep away from children.

SPECIFICATION

Item

Non-contact Infrared Scan function

Measurement Range

-60 to +500

 o

C (-76 to +932

 o

F)

Operating Range

0 to +50°C (32 to +122°F)

Accuracy (Tobj=15~35°C, Tamb=25°C)

±1.0°C (1.8°F)

A c c u r a c y  (Tobj=-33~500°C,
Tamb=23±3°C)

Tobj=0~500°C:

 ±2% of reading or 2°C (4°F) whichever is greater

Tobj=-60~0

 o

C: +/-(2°C+0.05/degree) °C

Emissivity Range

0.95 default – adjustable 0.1 to 1 step .01

Resolution (-9.9~199.9°C)

0.1°C/0.1°F

Response Time (90%)

1sec

Distance:Spot

12:1(90% energy covered)

Relative Humidity (Tamb=23+/-5degC)

1~99%, accuracy: +/-3% from 20~80%, ot/-5%

Dew point

-50~50°C, accuracy: ±2.5°C from 20~30%RH; ±2°C from 31~40%RH; ±1.5°C from
41~95 %RH

Battery Life

Typ.180, min 140 hours continuous use (Alkaline, without Laser and Back Light.)

Dimensions

46.0 x 143.0 x 184.8mm(1.81_5.63_7.28 inch)

Weight

240 grams(8.5 oz) including batteries (AAA*2pcs)

Note: Under an electromagnetic field of 3V/m from 545MHz to 679 MHz the maximum error is +/- 3.5_

 

EMC/RFI

Readings may be affected if the unit is operated within radio frequency electromagnetic field strength

of             

approximately 3 volts per meter, but the performance of the instrument will not be permanently affected.

U.S. PAT. B1 5,368,392; 5,524,984; 5,727,880; 5,823,678; 5,823,679; 6,123,453; 6,267,500 B1; 6,341,891 B1; 6,377,400 B1; 6,540,398 B2 ; 6,614,830 
B1; 6,633,434 B2; 6,659,639; 6,901,089 B1, 7,052,175; 7,307,255; 7,537,381; 7.611,278 / Canada 2,114,806; 2,223,195; 2,317,734 / France 2 756 920; 
2 767 921; 2,773,213; 2 773 214 / Germany G 94 22 197.9; G 94 22 203.7, 60024187; 69435097.4 / Holland 1007752; / U.K. Registered 2,320,324; 
9726133.3 / EPO 0 644,408 B2; EPO 1 085 307 A1, EPO 1 065 484 B1, EPO 1 065 483 B1. Other U.S. and International Patents Pending.

Reviews: