26
9.9
Automatic Emissivity Setting
This function is used to make an automatic setting of emissivity using an external thermocouple connected with the
auxiliary input.
•
If necessary, enable the thermocouple input (see par. 9.2) and select the thermocouple type (see par. 8.6)
•
Press the <
▲
> or <
▼
> key until the thermocouple temperature measure is displayed
e
°C
Lock
•
Press
<
ENTER
> key for setting the emissivity value with reference to the temperature values read from thermocouple
and pyrometer. The emissivity will change automatically to the correct value by the OSP.
e
°C
Lock
•
An error message appear if the
OSP
measure a temperature highest of the thermocouple. If the error happen, verify
the contact between the thermocouple and the object.
e
°C
Lock