33
•
Press the <
MEM
> key to start the acquisition.
If automatic acquisition mode is set,
OSP
acquires automatically with the programmed time interval.
If MANual mode is set, the real time temperature measurement is displayed; press the <
MEM
> key to store the
displayed value.
If MAX or MIN mode is set, the maximum (minimum) temperature measurement is displayed; press the <
MEM
> key
to trigger acquisition and the <
ENTER
> key to reset the value for the next acquisition.
•
Press the <
SEL
> key to obtain the page that allows the clearing of the memory stored data :
e
Log
°C
Lock
•
Press the <
ENTER
> key to clear the memory content and to obtain the following message that require confirmation :
e
Log
°C
Lock
•
Press the <
ENTER
> key to confirm the cancellation request and to obtain the following message :
e
Log
°C
Lock
•
Press the <
SEL
> key to obtain the following page indicating the identification code (or name) assigned to the active
Tag (zzzzzzz = Furn686 = Trap888). The identification code is selectable only through software.
e
Log
°C
Lock
•
Press the <
SEL
> key to review the pages relevant with -
HAL
- and -
LAL
- alarm settings. To modify the alarm level in
the data acquisition mode see the paragraph 10.2.