43
2 Remove the elevation and the windage adjustment caps.
3 Switch on the two pinpointing lasers and aim them to a target placed at a distance equal to the working distance.
4 The aiming must be carried out by keeping the eyepiece at a distance of about 80 cm from the eye.
5 Operating on the two adjustments (elevation and windage), place the center of the reticule in connection with the
center of the axis joining the two spot lasers. The minimum suggested working distance is 5 meters, shorter working
distance can cause alignment problems.
6 In this way the telescope is aligned to operate at the working distance; a variation in the working distance requires a
new alignment of the telescope.
Filter installation
1
Filter use is suggested in viewing targets with temperature higher than 1000°C
2
Install the filter ring on the front side of the telescope lightly pushing the component
CAUTION
NEVER LOOK AT THE SUN THROUGH THE TELESCOPE
:
THE CONCENTRATION OF STRONG SOLAR RAYS CAN CAUSE SERIOUS EYE
DAMAGE
.
NEVER LOOK AT TARGET WITH TEMPERATURE HIGHER THAN
1000°C
WITHOUT THE TELESCOPE PROTECTION FILTER
.
C
ONCENTRATION OF STRONG VISIBLE
- IR
RADIATION CAN CAUSE EYE DAMAGE
13.3
Red Point pinpointing system
Item
Q.ty
Part n.°
Description
1
2
EE010015
fastening screws for the guide
2
1
EE280169
guide
3
1
EE280180
N°1
red point sight device
N°2 mounting rings
N°1 polarize filter
N°1 lens Hood extension
4
1
EE360055
filter for T > 1300°C
5
1
EE360057
filter for T from 1000 to 1300°C
6
1
EE370127
O-ring/O-ring
Installation of the sighting device :
1. Remove with a cut screwdriver the nut protection caps placed on the upper part of the instrument.
2. Place the guide on the upper part of the pyrometer, taking care of orientating it so that it leans entirely on the back of
the instrument.
3. Fasten the screws in order to make the guide integral with the pyrometer.
4. Place the mounting rings on the guide, in order to cling them to the side edges. Let each ring slide along the guide
until each catch is housed in the respective slot. Fasten firmly each fastening screw with a coin or a screwdriver.
5. Separate the upper and the lower half of each mounting ring by removing the two screws, using the enclosed
hexagonal wrench.
6. Put the sighting device in the supporting framework formed by the lower halves of the mounting rings, so that the
rheostat hand grip is situated on the left side of the pyrometer, with respect to the aiming direction.
7. Rotate the sighting device into the mounting rings, in order to bring the elevation adjustment (UP) high. Fasten the
screws, avoiding to make an excessive clamping.
Alignment of the sighting device :
1. Make the pyrometer integral with a support.
2. Activate the illuminated red point by turning the rheostat hand grip clockwise to increase the brightness or in an
counter-clockwise direction to reduce it. The position “1” indicates the minimum brightness and the position “7”, the
maximum one.
3. The aiming must be carried out by keeping the eyepiece at a distance of at least 7 cm from the eye.
4. Remove the elevation and the windage adjustment caps.
5. Switch on the two pinpointing lasers and aim them to a target placed at a distance equal to the working distance.
The minimum suggested working distance is 5 meters, shorter working distances can cause alignment problems
6. Operating on the two adjustments (elevation and windage), place the illuminated red dot in connection with the
centre of the axis joining the two spot lasers.
7. In this way the sighting device is aligned to operate at the working distance ; a variation in the working distance
requires a new alignment of the sighting device.
Filter installation
1
Select the filter depending on the working temperature
2
Insert the selected filter in the internal side of the polarized filter mount
3
Place the O-ring over the filter to lock it