7
1.1
Instrument codes
To Order
Model No.
Temperature
Range
Spectral
Response
Target Spot Size
@ Distance **
D/S Optical Ratio
OSP500-(*)
–30 °C to +930 °C
-20°F to +1706°F
8 to 14
µ
m
10mm @ 600mm
0.39 @ 24”
60:1
OSP800-(*)
–30 °C to +930 °C
-20°F to +1610°F
8 to 14
µ
m
10mm @ 600mm
0.39 @ 24”
60:1
OSP1000-(*)
–30 °C to +1000 °C
-20°F to +1830°F
8 to 14
µ
m
10mm @ 1000mm
0.39 @ 39.4”
100:1
OSP1300-(*)
+300 °C to +1300 °C
+570°F to 2370°F
1.6
µ
m
6mm @ 1000mm
0.236 @ 39.4”
167:1
OSP1600-(*)
+600 °C to +1600 °C
+1120°F to 2900°F
1
µ
m
5mm @ 1000mm
0.20 @ 39.4”
200:1
OSP2000-(*)
+600 °C to +2000 °C
+1120°F to +3630°F
1
µ
m
5mm @ 1000mm
0.20 @ 39.4”
200:1
* Insert suffix codes B and C and D from options chart above
** See Target spot size diagram
Note
: all accessories and options must be specified at time of order.
Each Pyrometer supplied with
: Alkaline battery, laser, operators manual, and calibration report
Options
Suffix Code
Description
OSP500
OSP800
OSP1000
OSP1300
OSP1600
OSP2000
-B0
Alkaline battery
std
std
std
std
std
std
-B1
Rechargeable Battery
Opt
Opt
Opt
Opt
Opt
Opt
-C1
Single Laser
Std
No
No
No
No
No
-C2
Twin crossed Lasers
No
Std
Std
Std
Std
Std
-C3
Twin normal Lasers
No
Std
Std
std
Std
Std
-C4
Single Laser with Circle
Std
No
No
No
No
No
-D1
Vinyl Case
Std
Std
No
std
Std
No
-D2
Aluminum Case
Opt
Opt
Std
Opt
Opt
Std
-D3
SRS232 Cable
Opt
Opt
Std
Opt
Opt
Std
-D6
Sighting telescope
No
No
Opt
Opt
Opt
Opt
-D7
Red Point
No
No
No
Opt
Opt
Opt
-D8
Alarms
Std
Std
Std
std
std
Std
-D9
Analog Output
Opt
Std
std
std
std
Std
-DA
Aux Input for Tc K & S
Opt
std
std
std
std
Std
Report of Calibration
std
std
std
std
std
Std
No = Not available
Std = Standard at no extra cost
Single laser
: General application, target center
Single laser with circle
: General application, target center plus target dimension with circle (short distance)
Twin crossed lasers
: General application, target dimension
Twin normal lasers
(parallel): Long distance application over 5m (16’)
Red Point
: High temperature over 1250°C 1x
Sighting Telescope
: High temperature over 1250°C 2x tele
Software
(D3): Includes memory up to 500 values. You can divide them by Tag (20 different). Each value includes
temperature value and date&time (built in real time clock). Tag name with 7 characters on instrument
display. Datalogging manual or automatic with sampling time from 1 to 999 sec. Windows 95 software to
download data to PC and cable. You can view, save, print, export to Excel and graph data.
U.S. and international Patents and Patents Pending.