4.2.4.7 Recommended Measurement Parameters
If you are not sure which values to use for measuring parameters, the following values are recommended:
Emissivity: 0.90
Reflection Temperature: 20
℃
Environmental Temperature: 20
℃
Relative Humidity: 50%
Measuring Distance: 1.0m
External Optical Transmittance: 1.0
4.2.5 Temperature Measurement Range Selection
Click to select the temperature range to enter sub-menu.
! Note:
The most appropriate measurement range is the closest to the object temperature, which result a
more accurate measurement and better image quality.
4.2.6 Display Image Enlargement
Click for display image enlargement menu