background image

g.3

g.6

g.7

Connections tab

Logic board

X4

(There)

(B ')

(B)

(C)

(TO')

(TO)

(D)

CONNECTIONS

(6) - (7) After fixing the board, proceed with the connections (A, B, C, D) as shown in the figure.
(a) MPZ766 LAN cable
(b) MPZ756 FLAT cable

(c) MPZ1072 walking battery cable

(new cable included in the kit). Make sure 

that the cable does not interfere with mechanical components.

REMOVAL OF THE MPZ752C CONNECTION BOARD

(3) Gently disconnect the connections on the board and remove it from the "slot" by unscrewing the 4 

M3X6 screws

(including ground connection).

(d) grounding.

g.4

g.5

6

g.8

REPOSITIONING FRAME OF 

COVER AND CRUISE

Apply the upper closing frame, fixing it 

with the 10 M4X10 screws.

Then insert the spider, and fix it with 
the two 6X25 Allen screws previously 
removed.

Re-test the product.

INSERTION OF THE AMBULANT CONNECTION CARD cod. MPZ752F

(4) - (5) Insert the itinerant connection card cod. MPZ752F in the guides on the base, (before 
inserting the card it is recommended to connect the cable "C") and proceed with fixing, using 
the 4 screws, previously removed -restore the earthing in the same position.

2