background image

 

 

 

 
 
 
 

Omegasonics 

 

Ultrasonic Cleaning Equipment 

 

 
 
 

Operation & Instruction Manual  

 
 

OMEGASONICS 

Bench Top Unit 

Model OMG-1420BTD                 

 

 
 

120 VAC 15 A 

 
 
 
 
 

Read all instructions thoroughly before operating this equipment 

 

Summary of Contents for OMG-1420BTD

Page 1: ...1 Omegasonics Ultrasonic Cleaning Equipment Operation Instruction Manual OMEGASONICS Bench Top Unit Model OMG 1420BTD 120 VAC 15 A Read all instructions thoroughly before operating this equipment...

Page 2: ...ings 4 Set Up 5 General 5 Digital Temperature 6 Digital Timer 7 Cleaning Procedures 8 Machine Operation 8 FDM Parts 9 Equipment Maintenance 11 General 11 Tank Draining 11 Limited Warranty 12 Trouble S...

Page 3: ...d cleaning speed and consistency when compared with simple soaking or immersion with agitation Additionally the bubbles are small enough to penetrate even microscopic crevices cleaning them thoroughly...

Page 4: ...tainless steel welds Due to the heated liquid in the tank use baskets tongs or wires to insert or remove parts from the tank Do not operate the unit with wet hands Use only biodegradable cleaning agen...

Page 5: ...ncy Set digital timer and temperature controls The OMG 1420BTD utilizes 1 900 Watt built in silicone heating element and is well insulated The time required to heat the machine initially will vary bet...

Page 6: ...hysteresis which means that the heating mechanism and the OUT light will go on off based on a 5 variance This protects the life of the motor contactor AT ALM 2 Omegasonics use only Return Key Operatio...

Page 7: ...ay The ultrasound is factory set to 15 00 minutes To adjust the preset value press the blue keys Press the up arrow key to increase or down arrow key to decrease Left most blue key corresponds to 10 m...

Page 8: ...o the ON position The selector switch will spring back to the center position Before cleaning your first batch of items operate the ultrasound for fifteen 15 minutes This process is called degassing a...

Page 9: ...10 require 30 60 minutes c Larger parts 10 require 1 3 hours plus Activate the ultrasound by switching SONICS to ON Parts being cleaned do not require continuous supervision or labor intensive cleanin...

Page 10: ...SR100 SR110 ABS 158 F 158 F Polycarbonate 167 176 F Nylon 12 2 Part 145 F Nylon 12 3 Part 158 F PCABS 158 F 158 F Polycarbonate ISO 167 176 F ABS Translucent 158 F 158 F ABS I Medical 158 F 158 F Ulte...

Page 11: ...he inside of the tank with a clean soft cloth Do not use steel wool cleaning pads as they are too abrasive and will scratch the tank surface Rinse the tank again If the tank will not be used for a lon...

Page 12: ...e the responsibility of the user This warranty does not include damage or product failure which results from cavitation erosion misuse abuse or transportation damage This warranty is limited to the or...

Page 13: ...ook for the ice cube relay located at the left end of the back panel d Flip the window up window has a red button this bypasses the thermal overload You will see an orange flag visible in the upper ri...

Page 14: ...ill need to be replaced 7 Call Omegasonics to order a new circuit breaker Ultrasound circuit is not operational 1 Does the digital timer have a visible display 2 Does the timer count up 3 Does the tim...

Page 15: ...inspect the back panel wires for burnt or loose wires Burnt wires are typically caused by a loosened connection and must be replaced Thermal Overload Heat Sonics Circuit Ice Cube Relay Contactor Cont...

Page 16: ...Digital Timer Timer Digital Temperature Controller Digital Temp 3 Position Selector Sonics Switch 3 Position Selector 2 Position Selector Heat Switch 2 Position Selector Power Button Push Button Motor...

Reviews: