28
No lights on the control panel.
1. Is the Power push button pushed in?
2. Is there power to the facility outlet?
3. Turn off Power and unplug the machine.
4. Remove upper panel located on the right end of the machine.
5. Is the circuit breaker energized (UP-Red) or (DOWN-Green)?
6. If the circuit breaker is tripped, re-set the circuit breaker.
7. Plug in machine.
8.
Follow the steps above in “Isolate the exact issue”.
9. Machines manufactured after January 2015 have panel fuses and may be
burnt. If either of the fuse holder red lights are lit, the 5 amp fuse(s) must
be replaced.
Heat is causing the circuit breaker to trip.
1. Likely cause is a shorted heat blanket.
2. Remove the front (if applicable) and back lower access panels.
3. Inspect the heat blanket(s)
– they should be an orange/ pinkish color.
4. If the heat blankets are black or charcoal white, they must be replaced.
5. Call Omegasonics to order new heat blanket(s).
Pump is causing the circuit breaker to trip.
1. Likely cause is a faulty or shorted motor.
2. The timer could also be shorted, but the display will not display properly.
3. Call Omegasonics to order a new pump.
Ultrasound is causing the circuit breaker to trip.
1. Remove lower vented/ panel located on the right end of the machine.
2. Remove the power plug from the back of each generator.
3. Activate the ultrasound.
4. If the circuit breaker trips again, then the corrective action is to replace the digital
timer.
5. If the circuit breaker does not trip, then follow these procedures.
a. Plug in one generator at a time to isolate which exact generator is causing
issue.
b. Contact Omegasonics to have the existing generator repaired or replaced.
6. If no specific circuit causes the internal circuit breaker to trip but the issue occurs
sporadically, then the circuit breaker will need to be replaced.
7. Call Omegasonics to order a new circuit breaker.