background image

Reproducing the contents of this document in whole or in part for purposes other than private use is expressly prohibited in any form.  

Subject to changes and errors. All texts, images and marks are the property of NIMAX GmbH. 

 

33128_DE_Bedienungsanleitung_REV_A 

 

 

4 / 5 

2.5.

 

If you want to change the enlargement, turn the nosepiece 

(6) to change the lenses 4x - 100x (7).

 

Make sure that the lenses 

do not collide with the specimen when swivelling. It is advisable 
to first move the stage down slightly to create enough space 
between the lens and the specimen and then re-focus. A clear 
clicking sound confirms that the lens has been changed 
correctly. 

 

 

 

2.6.

 

When adjusting the focus, always ensure that the lens does not touch 

the specimen you are viewing. Therefore, move the coarse focus knob 
(11.1) only until the lens is three millimetres away from the coverslip of the 
specimen. Now slowly change the coarse adjustment until you can see a 
roughly sharp image and then use the fine adjustment (11.2) of the focusing 
to finally achieve image sharpness. If you now choose a higher 
enlargement, i.e. a different lens, you only have to turn the fine adjustment 
to finally achieve image sharpness again. 

2.7.

 

By turning the coaxial stage mechanism (12), the specimen can be 

manually moved precisely and carefully along the Y and Z axes through the 
eyepiece field of view.  

2.8.

 

The lateral fine adjustment, which moves the condenser from below towards or away from the specimen, 

can be used to control the degree of illumination and thus the contrast in the image. 

 

3. Maintenance and care

 

3.1. Maintenance To change the LED lamp, loosen the large slotted screw on the underside of the microscope 
and flip open the cover that is locked in place with it. You can now pull the defective LED lamp out of the 
socket, which is located on the inside of the cover, and replace it with a new LED lamp. Do not touch the new 
LED lamp with your hand. Fingerprints and dirt on the LED can affect its brightness. Clean the LED lamp with a 
clean, soft cloth. Handle the LED only with clean gloves or a piece of gauze and insert the new LED into the 
socket provided for the purpose. Close the cover again and re-tighten the previously mentioned slotted screw.

 

Caution: 

The bright LED lamp is very hot during operation and shortly after use. Before replacing the LED lamp or the 

fuse, be sure to disconnect the microscope from the power supply. 

 

 

 

 

11.1 
 
11.2 

Summary of Contents for BinoView 1000x

Page 1: ...rivate use is expressly prohibited in any form Subject to changes and errors All texts images and marks are the property of NIMAX GmbH 33128_DE_Bedienungsanleitung_REV_A 1 5 Operating manual image of the product Omegon BinoView Microscope 1000x English Version 03 2023 Rev A Item No 33128 ...

Page 2: ...transmitted light microscope Dimmable lighting LED Achromatic lenses 4x 10x 40x 100x Specimen protection for the 40x and 100x lens Work distances of the lenses o 4x 37 50 mm NA 0 10 o 10x 7 63 mm NA 0 25 o 40x 0 63 mm NA 0 65 o 100x 0 20 mm NA 1 25 Eyepieces WF 10x Magnifications 40x 100x 400x 1000x Oil Abbe condenser NA 1 25 with iris diaphragm Coaxial cross table Fine adjustment in X Y axis Coar...

Page 3: ...must not be damaged in any way to ensure safe use In the event of a defect contact the support department at NIMAX GmbH 2 Initial operation 2 1 Switch on the device and adjust the brightness on the dimmer wheel 16 to a pleasant but required luminous intensity 2 2 Adjust the pinhole on the condenser to change the light irradiation Adjust the design related aperture pinhole of the lenses to achieve ...

Page 4: ... only have to turn the fine adjustment to finally achieve image sharpness again 2 7 By turning the coaxial stage mechanism 12 the specimen can be manually moved precisely and carefully along the Y and Z axes through the eyepiece field of view 2 8 The lateral fine adjustment which moves the condenser from below towards or away from the specimen can be used to control the degree of illumination and ...

Page 5: ...ittle isopropanol alcohol pharmacy Please do not clean the lens surface too often with isopropanol Frequent use may cause scratches and damage to the coating resulting in poor light transmission and poor image quality 3 3 Care The microscope should be stored in a shady dry and clean place It must not come into contact with acids bases or vapours Do not expose the device to direct sunlight The work...

Reviews: