Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 11 of 21
Smart Measurement Solutions
Smart Measurement Solutions
®
Figure 25: noise floor / isolation
– measurement & trace settings
After that, start a single measurement.
Figure 26: dynamic range / isolation - start single sweep
Expected Measurement results:
Gain: max. - 100 dB for 1 Hz to 50 MHz
If you measure a gain higher than -100 dB please send in the Bode 100 for calibration / adjustment.
Figure 27: noise floor / isolation
– magnitude (dB) measurement