Bode 100 User Manual
46
4. Set:
• CH2: 50
Ω
ON (click on switch as shown)
• The switch
(before ATTN1) to the internal source as reference
Hint:
The following settings can be made either in the
Configuration
window or
in the
Configuration
area on the left side of the
Frequency Sweep
mode
window.
• ATTN 1 (channel 1 input attenuator): 10 dB
• ATTN 2 (channel 2 input attenuator): 10 dB
• Receiver bandwidth: 30 Hz
• Level: 0 dBm
Hint:
With a narrow receiver bandwidth like 30 Hz, the measurement is very
selective. Only little noise will affect the measurement and, consequently, the
measurements will be more stable but the sweep will be slow. The receiver
bandwidth of 3 kHz will perform the fastest sweep.