51
Frequency Sweep Mode
14.To find the parallel resonance frequency of the quartz filter, drag the cursor 2
to the lowest value (notch) of the upper graph.
In the marked area of the
Frequency Sweep
mode window, the series and
parallel resonance frequencies and the corresponding measurement data
are displayed.
Results
: Cursor 1 marks the series resonance frequency of 11.998 MHz and an
attenuation at the series resonance frequency of 1.47 dB.
Cursor 2 marks the parallel resonance frequency of 12.025 MHz and an
attenuation at the parallel resonance frequency of 80.31 dB.
For the measurement of the series resistance of the quartz filter we will use the
Smith chart. The Smith chart displays the reflection coefficient (see (Eq. 4-3) on
page 32) in the complex plane. The horizontal axis represents the real
component and the vertical axis the imaginary component of the DUT's
Summary of Contents for Bode 100