Bode 100 User Manual
60
The
probe calibration
will be
switched off automatically
if the following
parameters are changed:
• Reference mode (internal/external reference)
• Conversion ratio of external probes (see 7.4 "Using Probes" on page 92)
• Input resistance of the channel 1 and/or channel 2 (low/high impedance)
Hint:
Use the probe calibration if measurement parameters have to be changed
often during the measurements. You will save time because you do not need to
recalibrate the
Bode 100
each time you changed the parameters.
6.1.2 User
Calibration
The user calibration is the most accurate calibration method available with the
Bode 100
. The user calibration is performed at the exact measurement
frequencies. In the
Gain/Phase
and
Impedance/Reflection
measurement
modes, the
Bode 100
is calibrated at the source frequency. In the
Frequency
Sweep
mode, the calibration is performed at the exact frequencies specified by
the measurement points.
The user calibration allows
changing
the following
parameters without the
need of recalibrating
the
Bode 100
:
• Source level
• Attenuator 1 and attenuator 2
• Receiver bandwidth
• Zoom
without
the
Copy from Zoom
function (see "Copy from Zoom" on
The
user calibration
will be
switched off automatically
if one of the following
parameters is changed:
• Frequency values
• Sweep mode (linear/logarithmic)
• Number of measurement points (in the
Frequency Sweep
mode)
• Reference mode (internal/external reference)
• Conversion ratio of external probes (see 7.4 "Using Probes" on page 92)
• Input resistance of the channel 1 and/or channel 2 (low/high impedance)
• Zoom
with
the
Copy from Zoom
function (see "Copy from Zoom" on
Summary of Contents for Bode 100
Page 1: ...Bode 100 User Manual...
Page 6: ...Bode 100 User Manual 6 This page intentionally left blank...
Page 12: ...Bode 100 User Manual 12 This page intentionally left blank...
Page 14: ...Bode 100 User Manual 14 This page intentionally left blank...
Page 72: ...Bode 100 User Manual 72 This page intentionally left blank...
Page 81: ...81 Advanced Functions Figure 7 12 Measured curve with sweep settings copied from the zoom area...
Page 98: ...Bode 100 User Manual 98 This page intentionally left blank...
Page 100: ...Bode 100 User Manual 100 This page intentionally left blank...
Page 104: ...Bode 100 User Manual 104 This page intentionally left blank...