Test Set CMC 156
50
3 current outputs
1
range 1.25A
Output currents
3-phase AC (L-N)
1-phase AC (L-N)
in
parallel
1-phase DC (L-N)
3
×
0 to 1.25 A
1
×
0 to 1.25 A
1
×
0 to 3.75 A
1
×
0 to
±
1.25 A
Power
3-phase AC (L-N)
1-phase AC (L-L)
2
1-phase AC (L-N)
3
typ. 3
×
5 VA
typ. 1
×
10 VA
typ. 1
×
15 VA
guar. 3
×
4.9 VA
guar. 1
×
9.8 VA
guar. 1
×
14.7 VA
Resolution
< 50 µA
Accuracy
4
typical error
< 0.015 % rd.
+ 0.005 % of rg.
guaranteed error
< 0.04 % rd.
+ 0.01 % of rg.
Harmonic distortion
5
typ. 0.025 %
guar. < 0.05 %
Short-circuit protection
Unlimited against N and from L to L.
Open-circuit protection
Open outputs (idling) allowed.
Connection
4 mm banana sockets or amplifier
combination socket.
Isolation
Reinforced isolation to mains and to all SELV
interfaces.
Table 31
Current outputs, range 1.25 A
1
Data for 3-phase systems are valid under sym. conditions (0°, 120°, 240°).
2
Single-phase model (operation with two phases in phase opposition):
two currents in series, additional adapter SPA1 or SPA 156 recommended.
3
3 phases (3 x 1.25A) in parallel for in-phase operation.
4
rd. = reading; rg. = range, whereat
n
% of rg. means:
n
% of upper range value (1.25 A).
5
Values at 50/60 Hz, 20 kHz measuring band width, nominal value and nominal load.