Technical Data
41
6.3 Outputs
For block diagrams of the available generator outputs, please refer to
section 4.1, "Block Diagram" on page 20.
Table 6-3:
Analog current, voltage,
and LL outputs.
All voltages and current generators can independently be configured with
respect to amplitude, phase angle, and frequency.
All outputs are monitored. Overload conditions result in a message
displayed on the PC.
General Generator Outputs Data
(analog current and voltage outputs, outputs "LL out")
Frequency ranges
sinusoidal signals
1
harmonics / interharmonics
2
transient signals
1
Amplitude derating for current outputs at frequencies above 380 Hz.
2
Signals above 1 kHz are only supported in selected
Test Universe
modules and are only
available on the voltage outputs and the low level outputs.
10 … 1000 Hz
10 … 3000 Hz
DC … 3.1 kHz
Frequency resolution
< 5 µHz
Frequency accuracy
± 0.5 ppm
Frequency drift
± 1 ppm
Bandwidth (–3 dB)
3.1 kHz
Phase range
ϕ
- 360° to + 360°
Phase resolution
0.001°
Synchronized operation
Generator outputs can be synchronized
to a reference input signal on binary
input 10 (range: 40 … 70 Hz).
Temperature drift
0.0025 %/°C
Summary of Contents for CMC 353
Page 1: ...CMC 353 Reference Manual ...
Page 6: ...CMC 353 Reference Manual 6 ...
Page 18: ...CMC 353 Reference Manual 18 ...
Page 38: ...CMC 353 Reference Manual 38 ...
Page 70: ...CMC 353 Reference Manual 70 ...
Page 74: ...CMC 353 Reference Manual 74 ...
Page 90: ...CMC 353 Reference Manual 90 ...
Page 94: ...The OMICRON Bootloader Software Copyright Notices 94 ...
Page 96: ...Contact Information Technical Support 96 ...
Page 100: ...CMC 353 Reference Manual 100 ...