CMC 356 Reference Manual
68
6.6 Environmental
Conditions
6.6.1 Climate
Table 6-25:
Climate
6.6.2
Shock and Vibration
Table 6-26:
Shock and vibration
6.7 Mechanical
Data
Table 6-27:
Data regarding size and
weight
6.8 Cleaning
To clean the
CMC 356
, use a cloth dampened with isopropanol alcohol or
water. Prior to cleaning, always switch off the power switch and unplug the
power cord from the mains.
Climate
Operating temperature
0 ... +50 °C;
above +30 °C a 50 % duty cycle may apply.
Storage and transportation -25 … +70 °C
Max. altitude
2000 m
Humidity
5 … 95% relative humidity; no condensation
Climate
Tested according to IEC 60068-2-78
Dynamics
Vibration
Tested according to IEC 60068-2-6;
frequency range 10 ... 150 Hz; acceleration
2 g continuous (20 m/s²); 10 cycles per axis
Shock
Tested according to IEC 60068-2-27;
15 g / 11 ms, half-sinusoid, each axis
Size, Weight and Protection
Weight
16.8 kg (37 lbs)
Dimensions W x H x D (without handle)
450 x 145 x 390 mm
(17.7 x 5.7 x 15.4 ")
Housing
IP20 according to EN 60529