Preface
1 - 7
• If the power supply is
≤
190V AC,
CPC 100
cannot provide the full output
power at the
800A AC
output.
The same applies when an external current booster is used. Therefore, in
order to gain the full output power, provide a sufficient power supply
(190V ... 240V AC).
General
• Before connecting or disconnecting test objects and/or cables, turn off
CPC 100
by either the POWER ON/OFF switch or the Emergency Stop
button. Never connect or disconnect a test object while the outputs are
active.
• Make sure that a test object’s terminals that are to be connected to
CPC 100
do not carry any voltage potential. During a test, the only power source for a
test object may be
CPC 100
.
• At their output sockets and especially in the cables connected to them, in
operation the high-current
400A DC
and
800A AC
outputs generate a
significant amount of heat (approx. 300W/m at 800A). To prevent burns, use
gloves when touching the cables while in operation or a short while after.
• Do not insert objects (for example, screwdrivers, etc.) into any input/output
socket.
• Never use the test cards
Quick
and
Resistance
to measure the resistance
of windings with a high inductance because turning off the DC source results
in life-threatening voltage levels.
For this kind of measurement only use the special winding resistance test
card
RWinding
!
Caution:
The connector "Ext. Booster" is
always
galvanically connected to
mains, regardless whether or not an external booster is selected on
the software tab
Options | Device Setup
, the green warning
warning light (0) is on, the outputs are turned off or the Emergency
Stop button is pressed.
Handle with extreme caution. Do not use any other booster cables
than the ones supplied by OMICRON.
Note:
Even if you switched off
CPC 100
, wait until the red I/O warning
light is fully extinguished.
As long as this warning light is lit, there is still voltage and/or current
potential on one or more of the outputs.
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...