2 - 20
Starting the Software
When
CPC 100
is switched on, the operating system boots up, and the
CPC 100
software starts automatically.
During the hardware and software initialization phase, the software displays a
splash screen.
The
CPC 100
software starts in the Test Card View loading the
test procedure
default
(refer to ”The principles of test cards and test procedures” on page 17
Initially, the test procedure default contains one empty Quick test card
1
.
However, this default can be customized according to your specific
requirements with as many test cards as needed and of all types of your choice.
By creating a user-defined test procedure default,
CPC 100
starts up loading
your most frequently used test cards already containing specific values.
To learn more about defining such a default, refer to ”Creating Defaults and
Templates” on page 42 of this chapter.
1. More about Quick at ”Getting Started with Quick” in chapter 3.