5.1 Connecting VBO3 to a CMC 430
5.1.1 Max. 300 V output voltage
There is no star point at the
VBO3
voltage inputs. Usually, a star point is not necessary for use with a
CMC 430
test set. However, if you would like to use the
CMC 430
like one of OMICRON's other CMC
test sets and require an output voltage of only max. 300 V, you need to create a star point.
To create a star point:
•
Connect the
CMC 430
voltage outputs to one of the
VBO3
voltage input triples using a connection
cable with 4 mm safety banana plugs (included in the delivery).
•
Use flexible jumpers (included in the delivery) to interconnect the other
VBO3
voltage input triple,
and thus, create a star point.
•
Connect the star point to the
CMC 430
's neutral phase.
Flexible jumpers
1
2
3
N
Connection
cable
1
2
3
N
Test object
Connection cable
VBO3
CMC 430
VBO3 User Manual
12