5.2 Connecting VBO3 to other CMCs
There is no star point at the
VBO3
voltage inputs. Therefore, you need to create one when connecting
a CMC test set other than the
CMC 430
(i.e., the
CMC 256plus
,
CMC 310
,
CMC 353
, or
CMC 356
):
•
Connect the CMC voltage outputs to one of the
VBO3
voltage input triples using a connection
cable with 4 mm safety banana plugs (included in the delivery).
•
Use flexible jumpers (included in the delivery) to interconnect the other
VBO3
voltage input triple,
and thus, create a star point.
•
Connect the star point to the CMC's neutral phase.
Connection cable
VBO3
CMC 256plus
Test object
Flexible jumpers
Connection cable
1
2
3
N
1
2
3
N
VBO3 User Manual