8
SECTION 2 — MOTOR MAINTENANCE
2.1 MOTOR DRIVE UNIT
The Motor drive housing should be wiped off after use, especially
when concentrated and potentially damaging liquids are used during
processing. Never use solvents to clean unit or accessories.
2.2 BRUSH MAINTENANCE
If the performance of your Omni GLH begins to deteriorate in any of
the following ways:
•
The operations speed begins to decline
•
The unit stalls intermittently
•
The unit stops working all together
It probably requires a new set of brushes. The average brush life for
the Omni GLH is 150 hours. For service, or brush replacement, please
contact Omni International at 1-800-776-4431.
2.3 MOTOR BEARINGS
The Omni GLH motor is equipped with sealed ball bearings. Under
normal use they require no additional lubrication.
2.4 STORAGE
WARNING:
Keep all housings in place and in working order.
WARNING:
Remove all tools from the generator probe before turning
the motor on.
WARNING:
DO NOT
use the motor in a dangerous environment.
WARNING:
Disconnect the motor before servicing, and when chang-
ing the generator probe.