Operation instruction    

 

 

                     

 

 

                                                                                                 

1.Place compressor on a stable, strudy and flat surface, so that the unit can be easily reached when you are seated.
2.Pull the power cord from the base of the unit, make sure that the device is switched off (position O of the power switch ).
3.Plug power cord into the wall-socket.
4.Connect one end of the air tube into the air outlet connector on a nebulizer's body.
5.Unscrew the nebulizer cup (medicine container), check if the SPRAYING NOZZLE is inside. Pour the measure dose of 
medicine, prescribed by the doctor.
The sign 6 cc on the container points out the maximum container`s capacity. Do not pour more than  maximum level of 
medication solution. Tighten the container.

WITHOUT THE THE SPRAYING NOZZLE the NEBULIZER WILL NOT SPRAY A MEDICINE!

DO NOT LOSE THE COVER PLATE OF THE NOZZLE.

6.Put the mouthpiece or the mask on the medicine’s container (nebulizer cup).
7.Put the second end of the air tube on the bottom  part of the nebulizer cup, from the bottom. Sit down.
8.Switch on the device (position I of the power switch).
9.Put the mouthpiece between teeth, clench lips around the mouthpiece and breath deeply with a mouth.
10.When you use the mask, close attention to suit it to your face. Breath slowly and deepley, doing short breaks after 
each inspiration and doing fast expiration.

Inhalations should be conducted on the pitting position!
After finishing operations, the device should be switched off and taken out the plug from the socket.

Cleaning                                                                                                                                                  

1.

Turn the power off and unplug from the wall outlet .

2.

Remove the tubing from the air inlet.

3.

TO CLEAN:

 Disassemble mouthpiece, nosepiece, nebulizer cup, chamber and nozzle, and wash this items in hot

water with dish-washing detergent.Rinse these items thoroughly to remove the detergent and let air dry.

4.

TO DISINFECT: 

Mix one part of white vinegar with 3 parts hot water in a clean container. Submerge mouthpiece, 

nebulizer cap, chamber and baffle for half an hour in the solution. Remove from solution and let air dry.

5.

There is no need to clean the tubing. If necessary wipe the surface regularly.

WARNING: The nebulizer cup (medicine container) should be changed if it got dirty or plugged.

WARNING: Do not boil, do not scald out the medicine container and the other accessories.

It is recommended to replace accessories every 6 months or 3 months- if used intensively.

Maintenance        

 

 

                                                                                                                

 

1.

The air filter in the compressor should be changed if it only starts to darken. In order to do that raise up the filter’s 
cover, take out the used filter, put inside the new filter and shut the cover.Use only original filters and accessories!
Additional filters are available at the Seller or on the website: 

www.sklep-omnibus.pl

2.

Repairs can be made only by the authoised service. The service in Poland: Omnibus Medical Equipment, 9 
Grunwaldzka St., 82-500 Kwidzyn, 

www.omnibus-med.com.pl

WARNING: The nebulizer can be damaged, if it is used with the dirty or unoriginal filter.

DO NOT USE WITHOUT THE AIR FILTER.

Summary of Contents for Monkey MILO

Page 1: ...MONKEY MILO Monkey compressor piston nebulizer Manual CARREFULY READ MANUAL BEFORE FIRST USE Model nr BR CN001 Monkey MILO...

Page 2: ...IEC Symbols 3 Important Safeguards 3 Introduction 4 Specifications 4 Description 5 Accessories 5 Operation Instruction 6 Cleaning 6 Maintenance 6 Troubleshooting 7 Warranty 7 Technical data 8 Disposal...

Page 3: ...and other damage 1 The risk of electric shock Do not remove the cabinet or open the cover 2 Always disconnect the power cord from the electrical outlet before cleaning or servicing 3 Do not storage an...

Page 4: ...1 6 kg 3 53 lbs Noise level below 55 dB A Maximum compressor pressure 2 41 3 45 bar 241 345 kPa 35 Psi to 50 Psi Neb operating pressure 0 62 1 10 bar 62 110 kPa 9 16 Psi Maximum flow 7 11 l min Liter...

Page 5: ...on infant mask WARNING VERY IMPORTANT There should be close attention that without the spraying nozzle the nebulizer DOES NOT WORK Nebulizer cup consists of three parts upper part of cup lower part of...

Page 6: ...perations the device should be switched off and taken out the plug from the socket Cleaning 1 Turn the power off and unplug from the wall outlet 2 Remove the tubing from the air inlet 3 TO CLEAN Disas...

Page 7: ...re air vents holes air intake in the compressor body covered During operation don t cover air vents air intake in the compressor body If none of the solutions bought the result do not ty to repair the...

Page 8: ...ilized in accordance with the European Directive 2002 96 EC The device should be delivered to the special container or the specialized waste utilization point Marking with the emblem of the crossed wa...

Reviews: