• Cleaning the Nebulizer Kit, Mask, Mouthpiece, Nasal Shower and Adapter: 

Fully disassemble the Nebulizer Kit and the Nasal Shower. Wash them in warm water 
and mild, neutral detergent. Rinse them thoroughly with clean hot tap water, gently 
tap to remove excess water and allow to air dry in a clean place. It is advisable to 
replace the Nebulizer Kit and the Nasal shower after some 100 to 120 treatments or 
after about 20 boiling cycles. 
 

 

Cleaning the Compressor and Air Tube: 

Firstly, make sure that the power plug is unplugged from the power outlet. Wipe 
clean with a soft cloth moistened with water or mild, neutral detergent. 
 

 
Disinfecting the Nebulizer Kit, Mask, Nasal Shower, Mouthpiece and Adapter: 

Always disinfect the parts before using the product for the first time and after the 
product has not been used for an extended period of time as well as after the last 
treatment of the day. If the parts are heavily stained, replace them with new ones.  
The Nebulizer Kit, Mask, Nasal Shower, Mouthpiece and Adapter can be disinfected 
by 

 

using a chemical disinfectant

  such as ethanol, sodium hypochlorite (Milton), 

Quaternary ammonium (Osvan), Chlorhexidine (Hibitane) and Amphoteric Surfactant 
(Tego), following product instructions and rinsing off with clean warm water. Leave to 
air-dry. (Note: Never clean with benzene, thinner or a flammable chemical.) 

•  boiling:

  The Nebulizer Kit, Nasal Shower, Mouthpiece and Adapter can also be 

disinfected by boiling between 15 to 20 minutes in plenty of water. After boiling, 
carefully remove the parts, shake off excess water and allow to air-dry in a clean 
environment. (Note: Do not boil the Mask and the Air Tube.) 
 

Replacing the Air Filter: 

If the air filter has changed colour, or has been used for more than 70 applications, 
replace it with a new one. To purchase the filters – see “Other Optional/Replacement 
Parts” paragraph. Remove the air filter cover using a flat screwdriver as shown in 
figure; remove the filter and fit the new filter as shown. Put the air filter cover back in 
place. (Note:

 

Do not wash or clean the air filter. If the air filter Becomes wet, replace 

it. Damp air filters can cause blockages.) 
 
 

Icons in the IM:  
 

 

 
Class II equipment 

 

 
Alternating current 

 

 
Type BF applied part 

 

Serial number 

Summary of Contents for C102 Total

Page 1: ...0 23 55 44 701 www omron healthcare com BANKERS BANK OF TOKYO MITUSBISHI HOLLAND N V ACC NR 063 56 15 320 IBAN NL95BOTK0635615320 SWIFT BOTKNL2X CHAMBER OF COMMERCE NR 342 10306 BTW VAT NR NL8132 86...

Page 2: ...102 Total NE C102 E LEGAL MANUFACTURER PRODUCTION FACILITY 3A HEALTH CARE S r l Via Marziale Cerutti 90F G 25017 Lonato del Garda BS Italy DISTRIBUTOR OMRON HEALTHCARE EUROPE B V Scorpius 33 2132 LR H...

Page 3: ...itis etc The nasal shower relieves blocked nose and symptoms of conditions such as common cold allergic rhinitis or sinusitis This is a medical device Operate the device only as instructed by your doc...

Page 4: ...experts for home treatment The user should also be capable of understanding general operation of C102 Total and the content of the instruction manual Indications for use This product is intended to b...

Page 5: ...nd rhino pharyngeal cavities How to use the nebulizer 1 Make sure that the power switch is in the off position 2 Plug the power plug into a power outlet 3 Remove the inhalation top from the medication...

Page 6: ...ack the upper part 4 Connect one end of the air hose to the nasal shower and the other end to the compressor 5 Turn on the compressor to begin the treatment Operation 1 Operation is to be done by pres...

Page 7: ...100cm Mouthpiece Adapter Adult Mask PVC Child Mask PVC Instruction Manual Classifications Type BF Applied part Mouthpiece Nosepiece and Masks IP21 Ingress protection Nebulizer kit specifications Part...

Page 8: ...er AC Cord PVC polyvinyl chloride Rating label PET polyester Package Cardboard Instruction Manual Paper Materials used in accessories Item Material Nebulizer kit Mouthpiece Adapter PP polypropylene Ai...

Page 9: ...9 PM 1655 E 01 01 2018 Title page of Instruction Manual...

Page 10: ...10 PM 1655 E 01 01 2018 Unfolded package design of main device and accessories Main device...

Page 11: ...11 PM 1655 E 01 01 2018 Accessory set Model NEB ASKIT 11 Product Description Nebulizer Accessory Set...

Page 12: ...1 2013 Ed3 1 EN 60 601 1 2 2015 4th edition EN 60 601 1 11 2015 EN 60 601 1 6 2010 EN 62366 2008 EN 14971 2012 Operation and care Cleaning of the device Clean the parts after each use to remove residu...

Page 13: ...and Adapter can be disinfected by using a chemical disinfectant such as ethanol sodium hypochlorite Milton Quaternary ammonium Osvan Chlorhexidine Hibitane and Amphoteric Surfactant Tego following pr...

Page 14: ...s Product Waste Electrical Electronic Equipment Legal manufacturer Icons in the package Legal manufacturer This device fulfils the provisions of EC directive 93 42 EEC Medical Device Directive Read th...

Page 15: ...or respiratory specialist 10 This product should not be used on patients who are unconscious or are not breathing spontaneously 11 After finishing the treatment remember to disconnect the air tube fro...

Page 16: ...Pa Warranty and lifetime Warranty of the nebulizer 3 years Durable periods are as follows provided the product is used to nebulize 3 times a day for 10 minutes each time at room temperature 23 C Durab...

Reviews: