ES1-N
2
Ratings and Specifications
*1.
By default, the measurement range is 0 to 500°C, but you can change the range to -50 to +500°C or -50 to +1,000°C on the ES1-TOOLS.
*2.
Industrial electromagnetic environment (EN/IEC 61326-1 Table 2)
Measurement error is within ±10°C.
Type
Without laser pointer
With laser pointer
Item
Model
ES1-LP3-N
ES1-LP10-N
ES1-LW50-N
ES1-LW100-N
ES1-LW50L-N
ES1-LW100L-N
Detection element
Thermopile
Lens
Silicon
Measurement
wavelength
8 to 14 µm
Current
output
USB
output
Resolution: Approx. 0.1°C
Current
output
4 to 20 mA, -50 to 500°C
*
1
Load impedance: 250
Ω
max.
Resolution: Approx. 0.24
μ
A
4 to 20 mA, -50 to 1,000°C
*1
Load impedance: 250
Ω
max.
Resolution: Approx. 0.24
μ
A
4 to 20mA/-50 to 500°C
*1
Load impedance: 250
Ω
max.
Resolution: Approx. 0.24
μ
A
4 to 20 mA, -50 to 1,000°C
*1
Load impedance: 250
Ω
max.
Resolution: Approx. 0.24
μ
A
Power supply voltage
12 to 24 VDC/USB bus power (Functions other than current output can be used with USB bus power only)
Allowed power supply
voltage fluctuation
95% to 105% of the power supply voltage
Current consumption
30 mA max. (24 VDC)
• 30 mA max. (24 VDC, Not using laser pointer)
• 40 mA max. (24 VDC, Using laser pointer)
Cable length
2 m ±5 cm
Measurement accuracy
USB output
• Within ±(-8%rdg+1)°C
(Measurement temperature: -50 to 0°C)
• Within ±1°C
(0 to 200°C)
• Within ±0.5%rdg°C
(200 to 500°C)
Current output
• Difference from USB output is (±0.1% of output
range) °C or less
USB output
• Within ±(-8%rdg+1)°C
(Measurement
temperature: -50 to 0°C)
• Within ±1°C
(0 to 200°C)
• Within ±0.5%rdg°C
(200 to 1,000°C)
Current output
• Difference from USB
output is (±0.1% of
output range) °C or less
USB output
• Within ±(-8%rdg+1)°C
(Measurement
temperature: -50 to 0°C)
• Within ±1°C
(0 to 200°C)
• Within ±0.5%rdg°C
(200 to 500°C)
Current output
• Difference from USB
output is (±0.1% of
output range) °C or less
USB output
• Within ±(-8%rdg+1)°C
(Measurement
temperature: -50 to 0°C)
• Within ±1°C
(0 to 200°C)
Within ±0.5%rdg°C
(200 to 1,000°C)
Current output
• Difference from USB
output is (±0.1% of
output range) °C or less
Reproducibility
• Within 0.5°C
• Within ±1°C
(Measurement
temperature: -50 to 0°C)
• Within ±0.5°C
(0 to 500°C)
• Within ±1°C
(500 to 1,000°C)
• Within 0.5°C
• Within ±1°C
(Measurement
temperature: -50 to 0°C)
• Within ±0.5°C
(0 to 500°C)
• Within ±1°C
(500 to 1,000°C)
Temperature drift
• Within ±0.5°C/°C
(Measurement temperature: -50 to 0°C)
• Within ±0.25°C/°C
(Measurement temperature: 0 to 500°C)
• Within ±0.5°C/°C
(Measurement
temperature: -50 to 0°C)
• Within ±0.25°C/°C
(Measurement
temperature: 0 to 500°C)
• Within ±0.5°C/°C
(Measurement
temperature: 500 to
1,000°C)
• Within ±0.5°C/°C
(Measurement
temperature: -50 to 0°C)
• Within ±0.25°C/°C
(Measurement
temperature: 0 to
500°C)
• Within ±0.5°C/°C
(Measurement
temperature: -50 to 0°C)
• Within ±0.25°C/°C
(Measurement
temperature: 0 to 500°C)
• Within ±0.5°C/°C
(Measurement
temperature: 500 to
1,000°C)
Influence
of EMS
*
2
Radiated
electromag
netic field
immunity
Within ±10°C (80 MHz to 1.0 GHz)
Immunity
Conducted
Disturbance
Within ±10°C (150 kHz to 80 MHz)
Response time
Current output: 0.14 s max. (95% response, movement average of 1)
Emissivity setting
Factory setting: 0.95: Can be changed to a value between 0.100 and 1.999 with the setting tool
Moving average
function
Factory setting: 10
Can be changed to a value between 1 and 1,000
with the setting tool
Factory setting: 50
Can be changed to a value
between 1 and 1,000 with
the setting tool
Factory setting: 10
Can be changed to a value
between 1 and 1,000 with
the setting tool
Factory setting: 50
Can be changed to a value
between 1 and 1,000 with
the setting tool
Operating temperature
and humidity range
Temperature:0 to 55°C, Humidity:35 to 85% (without condensation)
Storage temperature
and humidity range
−
20 to 55°C (without condensation)
Vibration
resistance
Malfunction
10 to 55 Hz, 20 m/s
2
along 3 axes for 10 min.
Destruction
10 to 55 Hz, 0.75-mm single amplitude, along 3 axis for 2 hours
Degree of protection
None
Applicable safety
standards
CE, KC, FCC, RCM
-
IEC60825-1, PSC, FDA
Dimensions
L: 58, W: 32, H: 40 mm
L: 90.3, W: 32, H: 40 mm
L: 60.4, W: 32, H: 40 mm
L: 92.7, W: 32, H: 40 mm
Weight
Approx. 95 g
Approx. 115 g
Approx. 95 g
Approx. 115 g
Standard accessories
User’s Manual, Mounting
Brackets, Power/output cable (2
m), and Installation Gauge
User’s Manual, Mounting Brackets, and Power/output cable (2 m)