TROUBLESHOOTING
If this happens...
Possible causes...
Try this solution...
Pad gel does not stick
to skin.
Are you using pad
during perspiring?
Dry the pad placement
area.
Have the pads been
washed too long and/
or too frequently?
Leave the pad in
freezer for overnight.
Were the pads
stored under high
temperature, high
humidity, or direct
sunshine?
Replace both pads.
Pad Light is blinking.
Are both pads
attached to the body?
Re-attach dislocated
pad(s) onto the skin
firmly.
If this happens...
Possible causes...
Try this solution...
Pad Light is blinking.
Have you removed the
transparent film from
the pad?
Peel off film on the
adhesive surface of
pads.
Is the cord properly
connected to the main
unit?
Connect cord plug
correctly into the jack
at the bottom of the
main unit.
Is the adhesive
surface of pads dirty
or dry?
Wash adhesive
surface of pads softly
with your fingertips
for about 3 seconds
under slow running
cold water.
If the above measures are not effective, please consult your local
OMRON representatives.