Thank you for purchasing the OMRON JPN1 Automatic Blood Pressure 

Monitor.
The OMRON JPN1 is a compact, fully automatic blood pressure monitor, 

operating on the oscillometric principle. It measures your blood pressure and 

pulse rate simply and quickly. For comfortable controlled inflation without the 

need of pressure pre-setting or re-inflation the devices uses its advanced 

“IntelliSense” technology. 
Intended Use

This product is designed to measure the blood pressure and pulse rate of 

people within the range of the designated arm cuff, following the instructions in 

this instruction manual.

It is mainly designed for general household use. Please read the Important 

Safety Information in this instruction manual before using the unit.

Please read this instruction manual thoroughly before using the unit.

Please keep for future reference.

For specific information about your own blood pressure, CONSULT YOUR 

DOCTOR.

Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia 

or arteriosclerosis.

Please read this section carefully before using the unit.

Warning:   

• Indicates a potentially hazardous situation which, if not avoided, could result 

in death or serious injury.

(General Usage)

• Always consult your doctor. Self-diagnosis of measurement results and 

self-treatment are dangerous.

• People with severe blood flow problems, or blood disorders, should consult a 

doctor before using the unit, as cuff inflation can cause internal bleeding.

(AC Adapter Usage) 

 • Never plug in or unplug the power cord from the electric outlet with wet 

hands.

(Battery Usage)

• If battery fluid should get in your eyes, immediately rinse with plenty of clean 

water. Consult a doctor immediately.

Caution: 

• Indicates a potentially hazardous situation which, if not avoided, may result in 

minor or moderate injury to the user or patient or damage to the equipment or 

other property.

(General Usage)

• Do not leave the unit unattended with infants or persons who cannot express 

their consent.

• Do not use the unit for any purpose other than measuring blood pressure.

• Do not disassemble the unit or arm cuff.

• Use only the approved arm cuff for this unit. Use of other arm cuffs may result 

in incorrect measurement results.

• Make sure that the air tube is not wrapped around other parts of your body 

when taking measurements. This could result in injury when the air pressure 

in the air tube is increased.

• Do not leave the cuff wrapped on the arm if taking measurements during the 

night. This could result in injury.

• Do not inflate the arm cuff over 299 mmHg.

• Do not use a mobile phone or other devices that emit electromagnetic fields, 

near the unit. This may result in incorrect operation of the unit.

• Do not operate the unit in a moving vehicle (car, airplane).

• To inflate the cuff manually, refer to Section 3.3. If the cuff is over inflated, it 

can cause internal bleeding.

(AC Adapter Usage) 

• Use only the original AC adapter (optional) designed for this unit. Use of 

unsupported adapters may damage and/or may be hazardous to the unit.

• Plug the AC adapter into the appropriate voltage outlet. Do not use a 

multiple-tap. 

• Do not use the AC adapter if the unit or the power cord is damaged. Turn off 

the power and unplug the power cord immediately.

(Battery Usage)

• If battery fluid should get on your skin or clothing, immediately rinse with 

plenty of clean water.

• Use only four “AA” alkaline or manganese batteries with this unit. Do not use 

other types of batteries.

• Do not insert the batteries with their polarities incorrectly aligned.

• Replace old batteries with new ones immediately. Replace all four batteries at 

the same time.

• Remove the batteries if the unit will not be used for three months or more.

• Do not use new and used batteries together.

General Precautions

• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.

• Do not take measurements after bathing, drinking alcohol or caffeine, 

smoking, exercising or eating.

• Do not forcibly bend the arm cuff or bend the air tube excessively.

• When removing the air tube, pull on the air plug at the connection with the 

main unit not the tube itself.

• Do not inflate the arm cuff when it is not wrapped around your arm.

• Do not wash the arm cuff or immerse it in water.

• Read and follow the “Important information regarding Electro Magnetic 

Compatibility (EMC)” in the Technical Data Section.

• Read and follow the “Correct Disposal of This Product” in the Technical Data 

Section when disposing of the device and any used accessories or optional 

parts.

Important Safety Information

Note: Subject to technical modification without prior notice. 

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC’s and mobile 

(cellular) telephones, medical devices in use may be susceptible to 

electromagnetic interference from other devices. Electromagnetic interference 

may result in incorrect operation of the medical device and create a potentially 

unsafe situation.

Medical devices should also not interfere with other devices.

In order to regulate the requirements for EMC (Electro Magnetic Compatibility) 

with the aim to prevent unsafe product situations, the EN60601-1-2:2007 

standard has been implemented. This standard defines the levels of immunity to 

electromagnetic interferences as well as maximum levels of electromagnetic 

emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this 

EN60601-1-2:2007 standard for both immunity and emissions.

Nevertheless, special precautions need to be observed:

• Do not use mobile (cellular) telephones and other devices, which generate 

strong electrical or electromagnetic fields, near the medical device. This may 

result in incorrect operation of the unit and create a potentially unsafe situation. 

Recommendation is to keep a minimum distance of 7 m. Verify correct operation 

of the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at 

OMRON HEALTHCARE EUROPE at the address mentioned in this instruction 

manual.

Documentation is also available at www.omron-healthcare.com.

This marking shown on the product or its literature, indicates that it 

should not be disposed of, with other household wastes at the end 

of its working life. To prevent possible harm to the environment or 

human health from uncontrolled waste disposal, please separate 

this from other types of wastes and recycle it responsibly to 

promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this 

product, or their local government office, for details of where and how they can 

take this item for environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions 

of the purchase contract. This product should not be mixed with other commercial 

wastes for disposal. 

This product does not contain any hazardous substances.

Correct Disposal of This Product 

(Waste Electrical & Electronic Equipment)

Made in Japan

Manufacturer

EU-representative

Asia Pacific HQ

Production facility

Introduction

1.   Overview

Main unit: 

Arm cuff: 

M. Systolic blood pressure

N.  Diastolic blood pressure

O  Heartbeat symbol

  (Flashes during measurement)

P.  Memory symbol

Q. Deflation symbol

R.  Average value symbol

2.1   Installing/Replacing the Batteries

2.   Preparation

1.
2.

Remove the battery cover.

2.2   Setting the Buzzer/ Date and Time

1.

Press the      button.

2.

Set the monitor to the correct date and time before 

taking a measurement for the first time.

3.

Press the START/STOP button to store the setting.

Notes:

• If the batteries have been removed for 30 seconds or more, the date 

and time setting will need to be reset.

• If the date and time are not set, “-:--” appears during or after 

measurement.

Insert four “AA” batteries 

as indicated in the battery 

compartment and then 

replace the battery cover.

Notes:

• If the battery low symbol (     ) appears on the display, turn the unit off 

then replace all batteries at the same time.

• The measurement values continue to be stored in memory even after 

the batteries are replaced.

Disposal of used batteries should be carried out in accordance with the 

national regulations for the disposal of batteries.

The bottom edge of the cuff should be 1 to 2 cm above the elbow. 

Marker (arrow under tube) is centred on the middle of your inner arm. 

Close the fabric fastener FIRMLY.

3.   Using the Unit

4.   Troubleshooting and Maintenance

1.

Insert the air plug into the air jack.

2.

Put your arm through the cuff loop.

3.

Position the arm correctly.

This criteria is for home blood pressure measurement.

Systolic Blood Pressure 

Above 135 mmHg 

Diastolic Blood Pressure 

Above 85 mmHg

3.1   Applying the Arm Cuff

3.3   Taking a Reading

6.   Technical Data

5.   Optional Parts

4.1   The Icons and Error Messages

Blink 

Lit 

4.2   Troubleshooting

4.3   Maintenance

Problem

3.4   Using the Memory Function

Calibration and Service

To protect your unit from damage, please observe the following:

Cause

Remedy

Notes:

• When you take a measurement on the right 

arm, air tube will be at the side of your 

elbow. Be careful not to rest your arm on 

the air tube.

Notes:

• To cancel a measurement, press the START/STOP button to turn off 

the unit and to release the air in the arm cuff.

• Remain still while taking a measurement.

Press the START/STOP button.

The cuff will start to inflate automatically.

To delete all the values stored in memory

• Press the START/STOP button and repeat 

measurement.

• If the problem continues, try replacing the 

batteries with new ones.

If this still does not solve the problem, contact 

your local OMRON representative.

• Do not subject the main unit and the cuff to extreme 

temperatures, humidity, moisture or direct sunlight.

• Do not fold the cuff or tubing tightly.

• Do not inflate the arm cuff over 299 mmHg.

• Do not disassemble the unit.

• Do not subject the unit to strong shocks or vibrations 

(for example, dropping the unit on the floor).

• Do not use volatile liquids to clean the main unit. 

• Do not wash the arm cuff or immerse it in water.

• Do not use petrol, thinners or similar solvents to clean 

the arm cuff.

• Do not carry out repairs of any kind yourself. If a defect 

occurs, consult your local OMRON representative.

• The accuracy of this blood pressure monitor has been 

carefully tested and is designed for a long service life.

• It is generally recommended to have the unit inspected 

every two years to ensure correct functioning and 

accuracy. Please consult your local OMRON 

representative.

• The unit should be cleaned with a soft, dry cloth.

• Use a soft, moistened cloth and soap to clean the arm 

cuff.

Automatic Blood Pressure Monitor

JPN1

LCD Digital Display

Oscillometric method

Pressure: 0 to 299 mmHg 

Pulse: 40 to 180/min.

Pressure: ±3 mmHg

Pulse: ±5% of display reading

Fuzzy-logic controlled by electric pump

Automatic pressure release valve

60 Measurements with date and time

DC6V   4W

4 “AA” batteries 1.5V 

or AC adapter (optional, INPUT AC100-240V 

50/60Hz 0.12A)

Approx. 1500 measurements with new alkaline 

batteries at a room temperature of 23°C

        = Type B           

Internally powered ME equipment (When using only 

the batteries)

       = Class II ME equipment (Optional AC adapter)

+10 to +40°C / Maximum: 30 to 90% RH

-20 to +60°C / Maximum: 10 to 95% RH

/ 700-1060 hPa

Approx. 340g without batteries

Approx. 120g

Approx. 123 (w) mm × 85 (h) mm × 141(l) mm

Approx. 146 mm × 446 mm (Medium cuff: arm 

circumference 22 to 32 cm)

Nylon and polyester

Main unit, cuff, instruction manual, battery set

Remove tight-fitting clothing or tight rolled up sleeve from your 

upper arm.

Do not place the cuff over thick clothes.

3.2   How to Sit Correctly

To take a measurement, you need to be 

relaxed and comfortably seated, under 

comfortable room temperature. No eating, 

smoking or exercising 30 minutes before 

taking a measurement.

• Sit on a chair with your feet flat on the floor.

• Sit upright with your back straight.

• The cuff should be at the same level as your 

heart. 

1 to 2 cm

• The blood pressure can differ between the right arm and the left arm, 

and therefore also the measured blood pressure values can be 

different. OMRON recommends to always use the same arm for 

measurement. If the values between the two arms differ substantially, 

please check with your doctor which arm to use for your 

measurement.

1.

2.

Remove the arm cuff.

3.

Press the START/STOP button to turn off the monitor.

The monitor automatically stores the measurement in its 

memory.

It will automatically turn off after five minutes.

The monitor automatically stores the result up to 60 sets. It can also 

calculate an average reading based on the measurements from the last 

three readings taken within 10 minutes. If there are only two readings in 

memory for that period, the average will be based on two readings. If 

there is one reading in memory for that period, the average will be 

based on one reading.

Notes:

• If the memory is full, the monitor will delete the oldest readings.

• When viewing the reading taken without setting the date and time, 

“-:--” is displayed instead of the date and time.

Note:  

Wait 2-3 minutes before taking another blood pressure 

measurement. Waiting between readings allows the arteries to 

return to the condition prior to taking the blood pressure 

measurement.

Note:  You cannot partially delete the stored readings.

Note: Please check with your local OMRON representatives for the 

appropriate optional part models.

When the memory symbol (    ) appears, first press the Memory 

button. Then while holding it down, press the START/STOP button 

simultaneously for about 2-3 seconds.

Self-diagnosis of measured results and treatment are dangerous. 

Please follow the instructions of your doctor.

START

INFLATE

DEFLATE

END

COMPLETE

Warning:

• Recent research suggests that the following 

values can be used as a guide to high blood 

pressure for measurements taken at home.

Important:

Irregular heartbeats are 

detected.

Movement during 

measurement.

The batteries are low.

The batteries are 

exhausted. 

Cuff is under inflated.

Movement during 

measurement.

Air plug disconnected.

Arm cuff not applied 

correctly.

Clothing is interfering with 

the arm cuff.

Air is leaking from the arm 

cuff.

The arm cuff was inflated 

above 299 mmHg when 

inflating the cuff manually. 

Device error.

Remove the arm cuff. Wait 

2-3 minutes and then take 

another measurement. 

Repeat the steps in section 

3.3. If this error continues to 

appear, contact your doctor.

The reading is 

extremely low (or 

high).

Arm cuff deflates too 

soon.

Arm cuff pressure 

does not rise.

Cannot measure or 

readings are too low 

or too high.

Nothing happens 

when you press the 

buttons.

Other problems.

Arm cuff not applied 

correctly.

Movement or talking 

during measurement.

Clothing is interfering 

with the arm cuff.

The air tube is not 

securely connected 

into the main unit.

Air is leaking from the 

arm cuff.

The arm cuff is loose.

The arm cuff has not 

been inflated 

sufficiently.

The batteries are 

empty.

The batteries have 

been inserted 

incorrectly.

Apply the arm cuff 

correctly. 

Refer to section 3.1.

Remain still and do not 

talk during 

measurement.

Refer to section 3.3.

Remove any clothing 

interfering with the arm 

cuff.

Refer to section 3.2.

Make sure that the air 

tube is connected 

securely. 

Refer to section 3.1.

Replace the arm cuff 

with a new one. 

Refer to Chapter 5.

Apply the cuff correctly 

so that it is firmly 

wrapped around the arm. 

Refer to section 3.1.

Inflate the cuff so that it 

is 30 to 40 mmHg above 

your previous 

measurement result.

Refer to section 3.3.

Replace the batteries 

with new ones.

Refer to section 2.1.

Insert the batteries with 

the correct (+/-) polarity.

Refer to section 2.1.

You should replace them 

with new ones ahead of 

time.

Refer to section 2.1.

Carefully read and repeat 

the steps in sect

ion 3

.

3

You should replace them 

with new ones at once.

Refer to section 2.1.

Carefully read and repeat 

the steps listed under 

section 3.3.

Repeat measurement. 

Remain still and do not talk 

during measurement.

Refer to section 3.3.

Insert the air plug securely.

Refer to section 3.1.

Apply the arm cuff correctly. 

Refer to section 3.1.

Remove any clothing 

interfering with the arm cuff.

Refer to section 3.1.

Replace cuff with new one.

Refer to Chapter 5.

Do not inflate the arm cuff 

above 299 mmHg.

Refer to section 3.3.

Contact your local OMRON 

representative.

Cause

Remedy

Error Display

1.

Unplug the air tube from the air connector.

2.

Gently fold the air tube into the arm cuff.

4.4   Storage

Note:

Do not bend the air tube excessively. 

Do not store the unit in the following situations:

• If the unit is wet.

• Locations exposed to extreme temperatures, humidity, direct sunlight, 

dust or corrosive vapours.

• Locations exposed to vibrations, shocks or where it will be at an 

angle.

1.

2.

Using the Optional AC Adapter

Insert the AC adapter plug 

into the AC adapter jack on 

the rear side of the main 

unit.
Plug the AC adapter into an 

electrical outlet.

Medium Arm Cuff

Arm circumference 

22 - 32 cm

Large Arm Cuff

Arm circumference 

32 - 42 cm 

AC Adapter

Product Description

Model

Display

Measurement Method

Measurement Range

Accuracy

Inflation

Deflation

Memory

Rating 

Power Source

Battery Life

Applied Part

Protection Against 

Electric Shock

Operating 

temperature/ Humidity

Storage temperature/ 

Humidity/ Air pressure

Console Weight

Cuff Weight

Outer Dimensions

Cuff Dimensions

Cuff Material

Package Content

• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device 

Directive). 

• 

This blood pressure monitor is designed according to the European Standard 

EN1060, Non-invasive sphygmomanometers Part 1: General Requirements 

and Part 3: Supplementary requirements for electromechanical blood 

pressure measuring systems.

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, Kyoto

617-0002 JAPAN

OMRON HEALTHCARE EUROPE B.V.

Scorpius 33, 2132 LR Hoofddorp

THE NETHERLANDS

www.omron-healthcare.com

OMRON HEALTHCARE Co., Ltd.

Matsusaka Factory

Mie, JAPAN

OMRON HEALTHCARE SINGAPORE PTE LTD.

438A Alexandra Road, #05-05/08

Alexandra Technopark, Singapore 119967

www.omron-healthcare.com.sg

Automatic Blood 
Pressure Monitor

Model JPN1

Instruction Manual

A. Display

B.  START/STOP button

C.  Memory button

D.  Up/Down (          ) buttons

E.  Buzzer and Date/Time 

setting (    ) button

F.  Air Jack

G. Battery compartment

H.  AC adapter jack (for optional AC 

adapter)

I.  Arm cuff (Medium cuff: arm 

circumference 22-32cm)

J. Marker

K.  Air tube

L.  Air plug

S.  Movement error symbol

T.  Date/Time display

U.  Blood pressure level indicator

V.  Pulse display

W. 

Battery low symbol

X. 

Irregular heartbeat symbol

Push     or     buttons to 

change.
Push the      button to confirm.

Next setting appears.

1)

2)

If your systolic pressure is more than 

220 mmHg

After the cuff starts to inflate, press and hold the 

START/STOP button until the monitor inflates 30 to 40 mmHg 

higher than your expected systolic pressure.

Notes:

• The monitor will not inflate above 299 mmHg.

• Do not apply more pressure than necessary.

• Your blood pressure monitor includes an irregular 

heartbeat feature. Irregular heartbeats can influence 

the results of the measurement. The irregular 

heartbeat algorithm automatically determines if the 

measurement is usable or needs to be repeated. If the 

measurement results are affected by irregular 

heartbeats but the result is valid, the result is shown 

together with the irregular heartbeat symbol (    ). If the 

irregular heartbeats cause the measurement to be 

invalid, no result is shown. If the irregular heartbeat 

symbol (    ) is shown after you have taken a measurement, repeat 

the measurement. If the irregular heartbeat symbol (    ) is shown 

frequently, please make your doctor aware of it.

Average Value

Alternating 

date/time display

View the Average 

Value

• If the previous 

reading was taken 

without setting the 

date and time, the 

average value is not 

calculated.

View the Previous 

Reading

• The Memory number 

appears for a second 

before the pulse rate 

is displayed. The 

newest set is 

numbered “1”.

View the Readings 

Stored in Memory

Previous Reading

Display: 

Recent

Old

To disconnect the AC adapter, unplug the AC adapter from the 

electrical outlet first and then remove the AC adapter plug from the 

main unit.

1883264-4A

Note:  The irregular heartbeat symbol (    ) may also be displayed with error  

 messages.

Reviews: