11
Specifications
Name: OMRON
®
Instant Ear Thermometer
Model:
MC-509N
Measuring Range:
32.0 °C to 42.2 °C (89.6 °F to 108.0 °F)
Accuracy:
Laboratory: ±0.2 °C, 32.0 ~ 42.2 °C (±0.4 °F, 89.6 ~ 108.0 °F)
Operating
10.0 °C to 40.0 °C (50.0 °F to 104.0 °F)
Temperature:
Storage
-20.0 °C to 50.0 °C (-4.0 °F to 122.0 °F) / 15% to 95%
Temperature:
RH non condensing
Display:
L
iquid
C
rystal
D
isplay with resolution 0.1 °C (0.1 °F )
Response Time:
Approx. 1 second
Acoustic:
a. The unit is turned ON and ready for the measurement:
1 short beep sound.
b. Complete the measurement: 1 long beep sound.
c. System error or malfunction: 3 short beeps.
Memory:
Displays the last measured temperature for 3 seconds
when unit is turned on.
Automatic Switch-off: Approx. 1 minute after last measurement is completed.
Battery:
One 3 volt battery, CR2032 (approx. 1,500 measurements)
Low Battery Indication: Replace the battery when the low battery symbol
is shown on the LCD.
Dimensions:
6'' x 1
1
⁄
8
'' x 1
3
⁄
4
'' ; (H) 153mm (L) x 31mm (W) x 45mm
Weight:
1.9oz. / 53g (with battery), 1.8oz. / 50g (w/o battery)
Standards:
Complies with ASTM E-1965-98 standards.
Components:
Storage Case (with built in lens cover applicator)
20 Lens Covers: type MC-509NLC, one-3volt Lithium
Battery: type CR2032, and Instruction Manual
Replacement
USE ONLY OMRON
®
Lens Covers model:
Lens Covers:
MC-509NLC (40 ct.)