20
4ROUBLESHOOTING
4ROUBLE
0OSSIBLE CAUSE
#ORRECTIVE ACTION
The nebulization rate is
EXTREMELY LOW
Batteries are low (the orange
BATTERY LOW INDICATOR mASHES
Replace the batteries with new
ones. Refer to page 9.
The mesh is stained or clogged.
1. Clean the mesh by boiling. Refer
to page 15.
2. Replace the mesh cap with a new
one. Refer to page 10.
The green power
indicator does not turn
on and the unit does not
nebulize.
The batteries are low.
Replace the batteries with new
ones. Refer to page 9.
The capacity of rechargeable
batteries is low.
Charge the batteries with a
commercially available recharger.
The batteries are inserted in the
wrong direction.
Re-insert the batteries in the correct
direction. Refer to page 9.
The medication container is not
installed correctly to the main unit.
Remove the medication container
from the main unit and install
correctly. Refer to page 11.
The green power
indicator turns on
but the unit does not
nebulize.
The vibrator is not immersed in the
medication.
Tilt the main unit in order to immerse
the vibrator in the medication.
The medication container is not
filled with medication.
Fill with medication. Refer to page
11.
The mesh is broken.
Replace with a new one. Refer to
page 10.