181
Z550
Operation Manual
S
E
CT
IO
N 5
Usin
g the
Dow
n
loa
d
Fu
ncti
on
SECTION 5
System Settings
Using the Download Function
The controller acquires information from the sensor at startup. The startup time can be shortened by
downloading this information into the controller and storing it.
To replace the sensor for connection to the controller, perform the following procedure:
1.Turn off the download function.
2.Replace the sensor.
3.Download the new sensor information.
4.Turn on the download function.
If the sensor is replaced while the download function is on, measurements cannot be taken correctly with the
new sensor because the old sensor information (such as the measurement range) is still stored.
Also, perform the above procedure when a sensor of the same model is used to replace the existing sensor.
1.
Enter the System mode and select [Environment].
p.174
The setting screen appears.
2.
Change the setting of [Download].
3.
Select [End].
The setting will be registered and the screen display returns to
1
.
Option
Meaning
OFF
(default setting)
Information is acquired via the sensor connection at startup. The startup time will be longer
than when the ON function is used.
ON
The information stored inside the controller is acquired at startup. The startup time will be
shortened by comparison to when the OFF function is used.
Scn 0
ٕ
System
ٕ
Comm
Profile
Environment
Download
Initialize
Version
Settings
Sensor
Scene
Monitor
1st scene
Coordinate
Dec. pts
Download
End
Scn 0
ٕ
System
ٕ
:
:
:
:
:
:
:
:
Expert
Mirror
Console
Image
Scn 0
Individual
6 digits
OFF
operation.book Page 181 Monday, June 6, 2005 10:19 AM