76
S
e
ct
ion 5
DETAILE
D
S
E
TTINGS
ZJ-SD
User’s Manual
Section 5
Linear Output
Setting the second point (B)
11.
Use the same procedure as the first point to
set the correction value for the second point.
12.
Press the ENT key.
Confirming setting results
If linear output correction has been registered correctly,
the sub-display will show [OK].
If not, the display indicates [NG].
Check that the current (or voltage) values for the two
points are not the same and execute again.
operation2.book Page 76 Wednesday, November 29, 2006 2:56 PM