Standard Specifications
110
ZR-RX70 User’s Manual
Analog Input Measurement Accuracy
Common to ZR-RX70A
*1
Operating environment 23°C ± 5°C
• Left for at least 30 minutes after the power supply is turned on
• Filter Line
• GND connection
• Thermocouple used is T:0.32
φ
, other:0.65
φ
Item
Description
Voltage
±0.25% of F.S.
Temperature *1
Thermocouple
Measurement Temperature Range (°C) Measurement accuracy
R/S
0
≤
TS
≤
100
± 7.0°C
100 < TS
≤
300
± 5.0°C
R: 300 < TS
≤
1600
± (0.05% of rdg + 3.0°C)
S: 300 < TS
≤
1760
± (0.05% of rdg + 3.0°C)
B
400
≤
TS
≤
600
± 5.5°C
600 < TS
≤
1820
± (0.05% of rdg + 3.0°C)
K
–200
≤
TS
≤
–100
± (0.05% of rdg + 3.0°C)
–100 < TS
≤
1370
± (0.05% of rdg + 2.0°C)
E
–200
≤
TS
≤
–100
± (0.05% of rdg + 3.0°C)
–100 < TS
≤
800
± (0.05% of rdg + 2.0°C)
T
–200
≤
TS
≤
–100
± (0.1% of rdg + 2.5°C)
–100 < TS
≤
400
± (0.1% of rdg + 1.5°C)
J
–200
≤
TS
≤
–100
± 3.7°C
–100 < TS
≤
100
± 2.7°C
100 < TS
≤
1100
± (0.05% of rdg + 2.0°C)
N
0
≤
TS
≤
1300
± (0.1% of rdg + 2.0°C)
W
0
≤
TS
≤
2315
± (0.1% of rdg + 2.5°C)
Summary of Contents for ZR-RX70
Page 1: ...User s Manual Portable Multi Logger ZR RX70 Cat No Z283 E1 02...
Page 12: ...10 ZR RX70 User s Manual MEMO...
Page 16: ...14 ZR RX70 User s Manual MEMO...
Page 108: ...WEB Server Function 106 ZR RX70 User s Manual MEMO...
Page 118: ...External Dimensions 116 ZR RX70 User s Manual MEMO...
Page 119: ...5 APPENDIX APPENDIX Error Messages and Countermeasures 118 Index 121 Revision History 124...
Page 122: ...Error Messages and Countermeasures 120 ZR RX70 User s Manual MEMO...
Page 125: ...Index ZR RX70 User s Manual 123 5 APPENDIX MEMO...