Standard Specifications
ZR-RX70 User’s Manual
4
SPECIFICATIONS
111
Main Functions
Item
ZR-RX70A
Data record Sampling speeds *1 10, 20, 50, 100, 200, 500 µs
1, 2, 5, 10, 20, 50, 100, 200, 500 ms, 1, 2, 5, 10, 20, 30, 60 s, External
Memory capacity
Internal RAM: Approx. 64 MB SDRAM
Internal flash memory: Approx. 256 MB Flash Memory
USB memory: No limit (except each file must be within 2GB)
Memory contents
Setup conditions
Measured data
Screen copy
Save destination
specification*2
Internal RAM, internal memory or External USB memory
Setting of memory
used for data capture
Set the number of data capture points.
Setting range: 10 to 1000000 points
Setting unit: In steps of one point
Pre-trigger
0 to 100% (Set in steps of 10%)
Auto save function*3 ON or OFF setting
ON: Automatically saves the data in the internal RAM to the internal flash
memory or USB memory.
OFF: Only temporarily retains data in the internal RAM (The data is lost at power-off).
PC
Interface
Interface types
Ethernet (10BASE-T/100BASE-TX)
USB (USB 2.0 HIGH-SPEED)
Ethernet functions
Web server function: Displays ZR-RX70 screen image on Web browser and
operates ZR-RX70 from the Web brouser.
FTP server function: Transfers and deletes files from internal memory and USB
memory.
NTP client function: Corrects the time of internal clock.
USB functions
A connector: Connection to USB memory
B connector: Connection to PC ( Select if controlled by the PC software or USB
drive mode.)
Realtime data
transfer speed
1 ms to 60 s
Monitor
Display
5.7 inch TFT color LCD (QVGA: 320 × 240 dots)
Display settings
Waveform Digital screen
Waveform screen
Digital Calculation Display screen
X-Y display
Display languages
Japanese, English, Chinese, Korean, German, French
Backlight life
50,000 hrs (when brightness is down to 50%), depends on operation environment
Screensaver function OFF, ON (10, 30 sec, 1, 2, 5, 10, 30, 60 min)
Calculation
functions
EU (scaling function) Voltage value can be changed to any value (unit)
4 points can be set for each channel.
Statistical calculation
functions*4s
Types of statistical calculation : Average, Max, Min, Peak, RMS
Number of operations :
Maximum of 2 can be set simultaneously
(during data recording), Maximum of 5 can be
set simultaneously (during data replay)
Method :
Data between cursors specified (during data
replay)
Summary of Contents for ZR-RX70
Page 1: ...User s Manual Portable Multi Logger ZR RX70 Cat No Z283 E1 02...
Page 12: ...10 ZR RX70 User s Manual MEMO...
Page 16: ...14 ZR RX70 User s Manual MEMO...
Page 108: ...WEB Server Function 106 ZR RX70 User s Manual MEMO...
Page 118: ...External Dimensions 116 ZR RX70 User s Manual MEMO...
Page 119: ...5 APPENDIX APPENDIX Error Messages and Countermeasures 118 Index 121 Revision History 124...
Page 122: ...Error Messages and Countermeasures 120 ZR RX70 User s Manual MEMO...
Page 125: ...Index ZR RX70 User s Manual 123 5 APPENDIX MEMO...