Logic Alarm Cable Connection and Functions
26
ZR-RX70 User’s Manual
Logic Alarm Cable Connection and Functions
The logic alarm cable (ZR-XRL1: option) enables logic/pulse input, external trigger/sampling input, and alarm
signal output.
Connect the logic alarm cable (ZR-XRL1: option) to the external input/output terminal as shown below.
Logic/Pulse Specifications
Internal Equivalence Circuit
Item
Description
Number of input channels
4
Input voltage range
0 to +24V max. (single-ended ground input)
Threshold level
Approx. +2.5V
Hysteresis
Approx. 0.5 V (+2.5 to +3 V)
Switch between logic and pulse input.
Logic alarm cable (ZR-XRL1: option)
Note
Summary of Contents for ZR-RX70
Page 1: ...User s Manual Portable Multi Logger ZR RX70 Cat No Z283 E1 02...
Page 12: ...10 ZR RX70 User s Manual MEMO...
Page 16: ...14 ZR RX70 User s Manual MEMO...
Page 108: ...WEB Server Function 106 ZR RX70 User s Manual MEMO...
Page 118: ...External Dimensions 116 ZR RX70 User s Manual MEMO...
Page 119: ...5 APPENDIX APPENDIX Error Messages and Countermeasures 118 Index 121 Revision History 124...
Page 122: ...Error Messages and Countermeasures 120 ZR RX70 User s Manual MEMO...
Page 125: ...Index ZR RX70 User s Manual 123 5 APPENDIX MEMO...