Setting Menus
ZR-RX70 User’s Manual
3
SETTINGS AN
D
MEASUREMENT
71
(3)-3 Stop Side Source Settings
Specifies trigger conditions to stop data capture.
The actual timing conditions for data capture are related to the timer.
See also Sections “(3)-1 Timer Mode” p. 70, “(3)-2 Start Side Source Settings” p. 70, and “(3)-7 Timer
(3)-4 Pre-trigger
Specifies the percentage of data to be captured before a trigger is generated.
* Pre-trigger can be selected only when the data capture destination is the internal RAM and the start side
source setting is other than OFF.
(3)-5 Repeated Capturing
Sets the repeat function for repeated capturing.
This selection can be specified when the data capture destination is the internal RAM or the stop side source
setting is other than Off.
(3)-6 Repeat Interval
Specifies the interval from the start of one data capture process to the start of another data capture process if
the repeat function is used.
This selection can be specified when the following three conditions are all met:
• the Repeated capturing is On,
• the start side source setting is Off or Time, and
• the stop side source setting is set to Time, or the stop side source setting is set to Off and the data capture
destination is set to Internal RAM.
Selection
Description
Off
Stops capturing data unconditionally (when the [START/STOP] key is pressed or the
time specified for the timer function arrives).
Level
Stops capturing data when a specified level condition is met.
⇒
When Level is selected, set the condition for each channel.
See p.72.
External Input
Stops capturing data when an input signal is received from an external trigger terminal.
* A trigger is generated when the voltage changes from 5V (open) to 0V (short circuit
to GND).
Time
Stops capturing data after a specified length of time elapses after starting data capture.
⇒
When Time is selected, set the time after which you want to stop capturing data.
Selection
Description
Off
The repeat function is not used.
On
The repeat function is used.
After one data capture process ends, the next data capture process starts (or a trigger
is waited for if the start side source setting is other than Off).
If the pre-trigger is 30%
Start side trigger
Time available for data capture
30%
Summary of Contents for ZR-RX70
Page 1: ...User s Manual Portable Multi Logger ZR RX70 Cat No Z283 E1 02...
Page 12: ...10 ZR RX70 User s Manual MEMO...
Page 16: ...14 ZR RX70 User s Manual MEMO...
Page 108: ...WEB Server Function 106 ZR RX70 User s Manual MEMO...
Page 118: ...External Dimensions 116 ZR RX70 User s Manual MEMO...
Page 119: ...5 APPENDIX APPENDIX Error Messages and Countermeasures 118 Index 121 Revision History 124...
Page 122: ...Error Messages and Countermeasures 120 ZR RX70 User s Manual MEMO...
Page 125: ...Index ZR RX70 User s Manual 123 5 APPENDIX MEMO...