Firmware update
380
ZW User's Manual
10-2 Firmware update
For information on how to obtain the latest version of the firmware, please contact your OMRON sales
representative.
After obtaining the latest version of the firmware, follow the procedure below to update the firmware.
Do not turn OFF the power supply to the Sensor Controller during updating. The Sensor Controller would no longer
start up properly.
Use PC tools (Sysmac Studio) to update.
Explorer pane
: [Device Group] | [(Sensor Name)] (double click)
Edit pane
: [Tools] icon (
)
1
Click [Sensor setup] - [Update], and select
the update file.
The update file extension is ".BIN". When the file
is selected, the [FirmwareWriteInWindow]
popup is displayed.
The current firmware version is displayed at
[Sensor setup]. Be sure to check the version
before updating the firmware.
2
Click [Start the update.].
A message to confirm to start the update
appears.
When you click [Start the update.], if the "Different
format" message is displayed, this means that the
format information for the connected Sensor
Controller and the specified file do not match. If
this happen, never continue with the firmware
update. The Sensor Controller breaks down and
no longer starts up properly.
Important
Important
Important
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......