Processing Item Data List
386
ZW User's Manual
44
Frequency filter
2
Filter type
0
0: OFF
1: High pass filter
2: Low pass filter
3: Band pass filter
Yes
Yes
3
Cut-off frequency
1
1 to 999999 (1 div: 0.001)
Yes
Yes
4
Cut-off frequency (upper)
999999
1 to 999999 (1 div: 0.001)
Yes
Yes
5
Cut-off frequency (lower)
1
1 to 999999 (1 div: 0.001)
Yes
Yes
45
DIFFERENTIAL 2
Differential mode
0
0: OFF
1: ON
Yes
Yes
3
Number of differential cycles
1
1 to 5000 (ms)
Yes
Yes
46
Hold
2
Hold mode
0
0: OFF
1: Peak
2: Bottom
3: Peak to peak
4: Auto peak
5: Auto bottom
6: AUTO PEAK TO PEAK
7: Average
8: Sample
Yes
Yes
3
Trigger method
0
0: External
1: Self-up trigger
2: Self-down trigger
Yes
Yes
4
Trigger level
0
-999999999 to 999999999
(nm)
Yes
Yes
5
TRIGGER HYSTERESIS
0.05% of
measuring
range
0 to 999.999999 (mm)
Yes
Yes
6
Trigger delay time
1
1 to 5000 (ms)
Yes
Yes
7
Sampling time
100
1 to 5000 (ms)
Yes
Yes
8
Trigger delay mode
0
0: OFF
1: ON
Yes
Yes
47
Zero reset
5
Offset when a zero reset is
executed Offset
0
-999999999 to 999999999
(nm)
Yes
Yes
7
ZERO RESET MODE
0
0: Real
1: Hold
Yes
Yes
64
Zero reset execution enabled/
disabled (Status)
1
0: OFF
1: ON
Yes
Yes
49
Judgment
output
2
LOW threshold value
-25% of
measuring
range
-999999999 to 999999999
(nm)
Yes
Yes
3
HIGH threshold value
+25% of
measuring
range
-999999999 to 999999999
(nm)
Yes
Yes
Unit numbers 40 to 49 are parameters for the TASK 1 processing unit. To reference the parameters for the processing unit for TASK N, add 20 ×
(N - 1) to the unit number you want to reference.
(Example)
To change the average processing for TASK 2, reference the parameters for
Processing unit number = 43 + 20 × (2 - 1) = 63
Data number = 2
Unit
number
Processing Item Data
number
Parameter
default
value
Setting range/output range Acquisition
Yes/No
Setting
Yes/No
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......