Setting the Output Conditions
ZW User's Manual
87
4
Settings for
Func
tion
4-3 Setting the Output Conditions
Setting Scaling
This setting is used when you want to correct any errors that are generated due to the installation status of the
Sensor Head, and display the corrected value on the main display as a measured value.
There are three types of setting: "auto scaling" ("1-point scaling" and "2-point scaling") that automatically sets
the correction value of a placed sensing object, "fixed scaling" that manually sets the correction value and
"thickness scaling" that automatically sets the correction value of a placed sensing object (transparent object).
The "Zero reset" p.65 settings return to the default settings when scaling is set. If scaling is set, perform a zero reset
as necessary.
Performing Manual Scaling
Set the scaling by entering the correction value.
Item
Setting item
Setting value
Description
Scaling
(Manual scaling)
Scaling
OFF (default value)/ON
Set scaling ON/OFF.
Span
-2.0000 to 2.0000
(default value: 1.0000)
Set the inclination of the sensor characteristics as a
coefficient.
Offset
-999.999999 to 999.999999
(default value: 0)
A fixed value is added to/subtracted from the measured
value.
Important
-2.0
2.0
Measurement value
(mm)
Workpiece
displacement
Measurement value
(mm)
Workpiece
displacement
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......