Functions and Operations during Measurement
ZW User's Manual
301
Senso
r con
trol
ler ope
ra
ti
ons
9
Changing Threshold Values
The threshold value for the task displayed on the main display in the RUN mode can be changed.
To change the target task, change the task displayed on the main display.
Changing Display Details p.300
As an example, here is an explanation of the procedure for setting the HIGH threshold value by entering
"1.000" directly.
Operating procedure
Steps
Key operation
Display
Description
1
Press the
key in the RUN mode to display the setting
menu.
2
Press the
/
key to select H.JUDGE, then press the
key.
3
Set "1.000" as the HIGH threshold value.
The current measured value is displayed on the main display and the
current set threshold value is displayed on the sub-display. (At this
time, the THRESHOLD-H indicator lights up.)
Press
key to enter the editing mode, and the sub-display
blinks.
To edit numerical values, use the
/
/
/
keys.
Input "1.000", then press the
key.
4
Press the
key to leave the setting menu.
Important
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......