Installation
42
ZW User's Manual
Calibrating Sensor Head
Calibrate the Sensor Head by obtaining the dark data in the no-incoming light status.
• When removing and inserting a fiber cable from/to the Sensor Controller (Including the initial connection).
• When extending a fiber cable.
The calibration data is stored inside the Sensor Controller. Therefore, if the sensor head is calibrated once, as
long as the fiber connection state does not change (if it is not disconnected from or connected to the Sensor
Controller or an extension added), this operation is not needed each time.
The Sensor Head should be calibrated without any object within the measuring range or with the tip of the
Sensor Head shielded from light with an object with diffuse reflection.
(Correct calibration is not possible with a transparent object, semi-transparent object, or mirror. )
Explorer pane : [Device Group] | [(Sensor Name)] | [System] | [System Data] (double click)
Edit pane: [Sensor Settings] icon (
)
1
Set the operation mode to FUN mode.
3-3 Switching operation modes p.61
2
Click [Sensor head calibration] - [Execute
sensor head calibration].
Sensor head calibration is executed.
Sensor head
Measuring range
• Without any object
within the measuring range
• With light shielded
Install an object with diffuse
reflection (metal, paper, plastic, etc.)
than the measurement range.
Sensor head
Measuring range
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......