Setting the Output Conditions
90
ZW User's Manual
Performing 2-point Scaling
Measurement is performed at two positions, and offset values are set for those measured values.
Example: When correcting display values to match actual distances
Separate the two specified points by at least 1% of the rated measuring range for the connected Sensor Head.
For example, for the ZW-S40, the two measured points must be separated by at least 12 mm
0.01 = 0.12 mm as the
measuring range is 12 mm (± 6 mm).
Explorer pane
: [Bank] | [(Bank Data Name)] (double click)
Edit pane
: [Task Settings] icon (
)
Task Settings window : [Scaling]
Item
Setting item
Setting value
Description
Actual scaling
(two points)
Set value
(1st point)
-999.999999 to 999.999999 [mm]
Correct the measured value of the 1st point, and set the
value to display.
Set value
(2nd point)
-999.999999 to 999.999999 [mm]
Correct the measured value of the 2nd point, and set the
value to display.
1
Select [Scaling] ON.
2
Click [Auto setting].
The [Scaling] popup menu appears.
4.0
-3.0
-2.8
3.8
-3.0
4.0
Display value
(mm)
Measured value
(mm)
Point 2
Point 1
Important
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......