Setting Measurement Items
232
ZW User's Manual
Measuring the Thickness
As an example, here is an explanation of the procedure for measuring the thickness from the top surface of
glass on which specular reflection occurs as the 1st surface to the 2nd surface (as shown in the diagram on
Operating procedure
Steps
Key operation
Display
Description
1
Press and hold the
key for two seconds to enter the FUN
mode.
2
Set the material of the target to "MIRROR."
Setting the Material of the Target to Measure p.228
If the target object is a transparent object, such as glass, on which specular reflection occurs, set the material of the
measurement object to "MIRROR" to obtain more accurate measurement result.
3
Set the surface to "SUR.1ST."
3-1
Press the
/
key to select MEAS, then press the
key.
3-2
Press the
/
key to select TASK1, then press the
key.
3-3
Press the
/
keys to select "ITEM" and press the
key.
3-4
Press the
/
keys to select "THICK" and press the
key.
3-5
Select "SUR1."
Press the
/
keys to select "TOP" and press the
key.
3-6
Select the measurement surface.
EDGE1 to EDGE4: 1st surface to 4th surface
PEAK: Peak
The current setting value is displayed on the sub-display.
Press
/
key to enter editing mode and the sub-display blinks.
In this example, select "EDGE1" and press the
key.
4
Set the surface2 to "SUR.2ND."
Optional
Summary of Contents for ZW-C1*T
Page 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Page 116: ...Setting the System 114 ZW User s Manual MEMO ...
Page 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Page 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Page 319: ...Index ZW User s Manual 317 Index 11 ...
Page 321: ......