Setting the Output Conditions
96
ZW User's Manual
Hold clearing conditions
Held values can be cleared by the following operation.
• When FUN (adjustment) mode/RUN (operation) mode is switched
• When hold RESET is entered (20-pole terminal block,
key)
• Hold RESET command is entered.
(Example) Peak
(Example) Auto Peak
If an abnormal measured value is obtained during sampling, the hold value is cleared. To not clear the hold value even
if an abnormal measured value is obtained, set "KEEP" as the non-measurement setting.
Setting operation when measurement is not possible p.150
W
hen the hold
v
al
u
e is reset,
the clamp le
v
el is o
u
tp
u
t
u
ntil
the next hold
v
al
u
e is applied.
The mode is s
w
itched
b
et
w
een
the FU
N
mode and RU
N
mode
FU
N
mode
RU
N
mode
Sampling
Sampling
Hold
v
al
u
e
Meas
u
rement
v
al
u
e
Clamp le
v
el
W
hen RESET is exec
u
ted
(20-pole terminal
b
lock/key inp
u
t/command).
The mode is s
w
itched
b
et
w
een
the FU
N
mode and RU
N
mode
FU
N
mode
RU
N
mode
Hold
v
al
u
e
Meas
u
rement
v
al
u
e
Clamp le
v
el
W
hen RESET is exec
u
ted
(20-pole terminal
b
lock/key inp
u
t/command).
Important
Summary of Contents for ZW-C1*T
Page 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Page 116: ...Setting the System 114 ZW User s Manual MEMO ...
Page 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Page 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Page 319: ...Index ZW User s Manual 317 Index 11 ...
Page 321: ......