Installation
38
ZW User's Manual
Calibrating Sensor Head
Calibrate the Sensor Head by obtaining the dark data in the no-incoming light status.
• When removing and inserting a fiber cable from/to the Sensor Controller (Including the initial connection).
• When extending a fiber cable.
The calibration data is stored inside the
Sensor Controller
. Therefore, if the sensor head is calibrated once, as
long as the fiber connection state does not change (if it is not disconnected from or connected to the
Sensor
Controller
or an extension added), this operation is not needed each time.
The Sensor Head should be calibrated without any object within the measuring range or with the tip of the
Sensor Head shielded from light with an object with diffuse reflection.
(Correct calibration is not possible with a transparent object, semi-transparent object, or mirror. )
Explorer pane : [Device Group] | [(Sensor Name)] | [System] | [System Data] (double click)
Edit pane: [Sensor Settings] icon (
)
1
Set the operation mode to FUN mode.
3-3 Switching operation modes p.57
2
Click [Sensor head calibration] - [Execute
sensor head calibration].
Sensor head calibration is executed.
Sensor head
Measuring range
• Without any object
within the measuring range
• With light shielded
Install an object with diffuse
reflection (metal, paper, plastic, etc.)
than the measurement range.
Sensor head
Measuring range
Summary of Contents for ZW-C1*T
Page 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Page 116: ...Setting the System 114 ZW User s Manual MEMO ...
Page 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Page 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Page 319: ...Index ZW User s Manual 317 Index 11 ...
Page 321: ......