Reflective Sensor Heads: RUN Mode Functions
Section 3-3
38
3-3
Reflective Sensor Heads:
RUN Mode Functions
3-3-1
Sub-display Changes
Items shown on the sub-display can be selected.
When Intensity Mode is OFF, the threshold values (HIGH/LOW), voltage
value, current value, incident level, or resolution can be selected.
When Intensity Mode is ON, the threshold values (HIGH/LOW), voltage
value, current value, or resolution can be selected.
3-3-2
Zero Reset/Release
The following are performed for the zero reset function:
• Setting the display value to 0.
• Setting the linear output to the center output value between two points
set for the monitor focus when 0 is displayed (default current output: 12
mA, default voltage output: 0 V).
The zero reset can be also released.
Refer
to
4-3-4 Zero Reset Function
.
• Voltage display
ëëë
The voltage level of the linear output is dis-
played.
• Current display
ëëë
The current level of the linear output is dis-
played.
• Incident level display
ëëë
The incident level is displayed (0 to 9999)
• Resolution display
ëëë
The resolution of linear output is displayed.
Display values are provided as reference values. There may be
some discrepancies in actual outputs.
The incident level displayed here is different from the one displayed
on the main display when Intensity Mode is ON.
POINT
The inclination of the linear output value against the actual distance
does not change when the zero reset is executed.
An error will occur if the zero reset is performed outside the mea-
surement range.
POINT
Summary of Contents for ZX Series
Page 1: ...Cat No Z157 E1 01B OPERATION MANUAL ZX Series Smart Sensors ...
Page 2: ......
Page 3: ...Smart Sensors ZX Series Operation Manual Cat No Z157 E1 01B OMRON Corporation ...
Page 48: ...OutlineofFunctions Section 2 5 28 Memo ...
Page 160: ...FUN Function Mode Section 4 5 140 Memo ...
Page 166: ...Setting Problems Section 5 2 146 Memo ...
Page 186: ...Dimensions Section 6 2 166 Memo ...
Page 187: ......
Page 188: ...Z157 E1 01B ...