Detailed information
about your system
9
134
Precision
Within Run Precision (300 Venous Blood Samples Tested
per Glucose Level)
Data generated using the OneTouch Select® Plus Meter.
Target
Glucose
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coefficient
of Variation
(%)
2.2
2.50
0.09
3.59
3.6
3.90
0.10
2.56
6.7
7.06
0.16
2.25
11.1
11.57
0.29
2.52
19.4
19.55
0.40
2.07
Results show that the greatest variability observed
between test strips when tested with blood is 0.10 mmol/L
SD or less at glucose levels less than 5.55 mmol/L, or
2.52% CV or less at glucose levels at 5.55 mmol/L or
above.