Detailed information about your system
9
Regression Statistics
Samples were tested in duplicate on each of three test
strip lots. Results indicate that the OneTouch® Verio™
Pro System compares well with a laboratory method.
# of
Subjects
# of
Tests
Slope
Intercept
(mmol/L)
100
600
0.949
0.070
95% CI
Slope
95% CI
Intercept
Std. Error
(S
y.x
)
R
2
0.941 to
0.957
-0.022 to
0.163
0.615
0.990
System Accuracy Results across the
entire Glucose Range
Percent (and number) of meter results that match the
laboratory test
Within ±0.67 mmol/L or ±15%
99.0% (594/600)