88
Control solution testing
5
6. Cleaning
Clean the top of the control solution cap with a clean,
damp tissue or cloth.
Out-of-range results may be due to:
• Not following the instructions detailed on pages 82–87.
• Control solution is contaminated, expired, or past its
discard date.
• Test strip or test strip vial is damaged, expired, or past its
discard date.
• Meter, test strips and/or control solution were not all at
the same temperature when the control solution test
was performed.
• A problem with the meter.
• Dirt or contamination in the small well on the top of the
control solution cap (see Step 2).