3.5.5 Remote Mount Wiring
Remote Sensor Wiring Detail
INSIDE BODY VIEW
INSIDE REAR
COVER LABEL
A1 - RED - VELOCITY SENSOR WIRE (HEATED ELEMENT)
A2 - GREEN - SENSE WIRE
A3 - BLUE - VELOCITY SENSOR WIRE (HEATED ELEMENT)
A4 - WHITE - TEMPERATURE SENSOR WIRE
A5 - BLACK SENSE WIRE
A6 - ORANGE - TEMPERATURE SENSOR WIRE
B1 - AC1 - AC VOLTAGE
B2 - AC2 - AC VOLTAGE
B3 - NU - NOT USED
B4 - 4 - 20 mA POWER
B5 - VDC IN - VOLTAGE DC - POSITIVE (+)
B6 - VDC GND - VOLTAGE DC - GROUND (-)
C1 - NOT USED
C2 - RS 485 (+)
C3 - RS 485 (-)
C4 - 24 VDC PULSE - 0 TO 24 DC PULSE OUTPUT
C5 - 4 - 20 mA - 4 TO 20 mA ANALOG OUTPUT
C6 - VDC GND - VOLTAGE DC - GROUND (-)
½” NPT
USER ENTRY
FOR WIRING
(ONE ON EACH SIDE)
FLOW
½” NPT
USER ENTRY
FOR WIRING
1
2
3
4
5
6
1
2
3
4
5
6
A
B
C
1
RED
AC1
COM
1
2
GREEN AC2
B +
2
3
BLUE NU
A –
3
4
WHITE NU
4
5
BLACK
4-20mA 5
6
ORANGE
SIG GND 6
A
B
C
24 VDC
PULSE
VDC IN
+
VDC GND
–
REMOTE
TERMINAL BLOCK SIDE
NO CUSTOMER ACCESS
DO NOT OPEN THIS SIDE
JUNCTION BOX
¾” NPT FOR
REMOTE CABLE
¾” NPT FOR
REMOTE CABLE
JUNCTION BOX
CONTAINS NO
ELECTRONICS,
JUST TERMINALS
¾” NPT FOR
REMOTE CABLE
DISPLAY SIDE
*
*
Factory installed jumper and resistor. Do not remove.
*
Remote Cable
Blue
Green
Red
Orange
Black
White
Ground Screw
Metal conduit
is recommended
with appropriate
grounding to
minimize effects
from external
noise sources.
Interconnect Cable
¾” NPT
conduit connection
To Terminal A1 to A6
shown above
| sales@mvandc.com | Phone: 877.566.3837 | Fax: 925.407.2903