ZERO FLOW CALIBRATION TEST
The zero flow calibration test provides a simple, yet effective, method for periodic field validation
of the factory calibration. Prior to shipment to the customer, each ONICON F-5000 series flow meter
undergoes a multipoint calibration. In this process, signal levels are recorded at a series of known
velocities over the specified operating range. The last data point taken in this process is taken at zero
flow (zero velocity). The value recorded at zero flow is part of the calibration record provided with the
meter and is stamped on the meter nameplate. By utilizing the mW output displayed by the meter, it is
possible to duplicate the factory no flow condition (zero velocity) test in the field.
Successfully duplicating this test provides field validation of the current calibration. A successful
zero flow test confirms that the flow sensor circuitry operating characteristics have not changed and
validates that the zero velocity data point has not shifted or drifted. It also verifies that the sensor
assembly is clean and undamaged and does not require visual inspection. It essentially provides a
means to validate that the meter is operating properly and can be used as part of a quality assurance/
preventative maintenance program or in lieu of scheduled annual factory calibrations.
IMPORTANT NOTE
The original factory zero flow calibration was performed with a vertically oriented meter installed
through a valve located on the top of a horizontal test pipe. If the meter is installed in a different
orientation the output at zero flow will be slightly different.
Depending on the orientation, meters inserted horizontally into a pipe may increase the reading by
as much as 5-6 mW.
i
There are 3 ways to validate the meter’s zero flow calibration.
WARNING
SYSTEM MAY BE UNDER HIGH PRESSURE. When removing the flow meter, be sure to grasp
the electronics enclosure firmly by hand before slowly loosening the compression fitting. Failure
to do this will allow the pressure in the pipe to suddenly and rapidly force the meter from the
pipe causing serious injury. The hand effort required to hold the meter will be 0.2 times the pipe
pressure.
!
!
WARNING
Do not attempt to remove the meter from a pipe pressurized to greater than 250 PSIG. For pipes
pressurized above 250 PSIG isolate flow and relieve pressure in the pipe before attempting to
remove the meter.
Method 1
1. Close valves in the process supply line to stop all flow in the pipe where the meter is located.
2. Wait up to 2 – 3 minutes to allow the meter reading to stabilize at zero and note the mW
reading on the display.
3. Compare the reading to that stamped on the nameplate. The two readings should agree within
± 5mW.
| sales@mvandc.com | Phone: 877.566.3837 | Fax: 925.407.2903